A vascular stent as an active component for locally enhanced magnetic resonance imaging: initial in vivo imaging results after catheter-guided placement in rabbits.

Dietmar Kivelitz1, Susanne Wagner, Jorg Schnorr, Rainer Wetzler, Martin Busch, Andreas Melzer, Matthias Taupitz, Bernd Hamm.   

Abstract

RATIONALE AND
OBJECTIVE: A vascular stent constructed as a high frequency resonator improves the local signal-to-noise ratio at magnetic resonance (MR) imaging. After catheter placement and intravascular expansion, the stent can be used as an inductively coupled coil for MRI. The imaging properties of this balloon-expandable active MRI stent (AMRIS) were evaluated after x-ray fluoroscopy guided placement in the abdominal aorta of five rabbits using MR angiography (MRA) and flow measurements.
METHODS: The AMRIS was implanted in the abdominal aorta of five rabbits using a balloon catheter inserted through the common carotid artery. The rabbits were examined by MRA (3D fast low-angle shot) at 1.5 tesla before and after intravenous injection of an iron-oxide-based blood pool contrast medium (dose 50 micro mol Fe/kg) and flow measurements (ECG-triggered phase contrast cine gradient-echo sequence). Signal-to-noise ratios (SNR) were calculated and flow volume curves were generated. The in-stent increase in temperature was measured in vitro using a fiberoptic thermometry system.
RESULTS: The SNR was 5.0 +/- 0.6 outside the stent and 23.2 +/- 14.1 within the stent ( < 0.0 5) in plain MRA, 19.5 +/- 5.0 outside and 30.7 +/- 8.2 within the stent ( < 0.05) in contrast enhanced MRA, and 5.8 +/- 1.6 and 13.9 +/- 5.9, respectively ( < 0.05) in the magnitude images of the flow measurements. Flow volume curves within and distal to the stent were comparable.
CONCLUSIONS: The expandable active MRI stent produces local signal enhancement in MRA and MR flow measurements after catheter placement and thus may improve assessment of the stented vessel segment by MR imaging.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12595794     DOI: 10.1097/01.RLI.0000052981.82153.A1

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  8 in total

Review 1.  Cardiovascular interventional magnetic resonance imaging.

Authors:  Robert J Lederman
Journal:  Circulation       Date:  2005-11-08       Impact factor: 29.690

Review 2.  Real-time, Interactive MRI for cardiovascular interventions.

Authors:  Elliot R McVeigh; Michael A Guttman; Peter Kellman; Amish N Raval; Robert J Lederman
Journal:  Acad Radiol       Date:  2005-09       Impact factor: 3.173

3.  Initial feasibility testing of limited field of view magnetic resonance thermometry using a local cardiac radiofrequency coil.

Authors:  Nelly A Volland; Eugene G Kholmovski; Dennis L Parker; J Rock Hadley
Journal:  Magn Reson Med       Date:  2012-11-19       Impact factor: 4.668

Review 4.  Magnetic resonance imaging-guided vascular interventions.

Authors:  Cengizhan Ozturk; Michael Guttman; Elliot R McVeigh; Robert J Lederman
Journal:  Top Magn Reson Imaging       Date:  2005-10

5.  Wireless amplified NMR detector for improved visibility of image contrast in heterogeneous lesions.

Authors:  Xianchun Zeng; Shengqiang Xu; Changyong Cao; Jian Wang; Chunqi Qian
Journal:  NMR Biomed       Date:  2018-07-16       Impact factor: 4.044

6.  Finite volume analysis of temperature effects induced by active MRI implants: 2. Defects on active MRI implants causing hot spots.

Authors:  Martin H J Busch; Wolfgang Vollmann; Dietrich H W Grönemeyer
Journal:  Biomed Eng Online       Date:  2006-05-26       Impact factor: 2.819

7.  Finite volume analysis of temperature effects induced by active MRI implants with cylindrical symmetry: 1. Properly working devices.

Authors:  Martin H J Busch; Wolfgang Vollmann; Jörg Schnorr; Dietrich H W Grönemeyer
Journal:  Biomed Eng Online       Date:  2005-04-08       Impact factor: 2.819

8.  Wireless Reconfigurable RF Detector Array for Focal and Multiregional Signal Enhancement.

Authors:  Wei Qian; Xin Yu; Chunqi Qian
Journal:  IEEE Access       Date:  2020-07-24       Impact factor: 3.367

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.