Literature DB >> 12578072

Accuracy of capnography with a 30 foot nasal cannula for monitoring respiratory rate and end-tidal CO2 in children.

K P Mason1, P E Burrows, M M Dorsey, D Zurakowski, B Krauss.   

Abstract

We tested the accuracy of a low flow (50 cc/min) sidestream capnography system equipped with an experimental 30-foot nasal cannula to monitor ventilatory status in children. End-tidal CO2 and respiratory rate, both at room air and in the presence of supplemental oxygen, were recorded simultaneously from the experimental 30-foot nasal cannula and the standard, FDA approved, 10-foot nasal cannula. The 30-foot nasal cannula was as accurate as the 10-foot nasal cannula in measuring respiratory rate and end-tidal CO2 in children. When supplemental oxygen was delivered by face-mask, there was no dilutional effect on the respiratory rate or end-tidal CO2 recorded with either the 10-foot or 30-foot nasal cannulas in place.

Entities:  

Mesh:

Year:  2000        PMID: 12578072     DOI: 10.1023/a:1011436329848

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  6 in total

1.  Microstream capnograpy technology: a new approach to an old problem.

Authors:  Y Colman; B Krauss
Journal:  J Clin Monit Comput       Date:  1999-08       Impact factor: 2.502

2.  Capnometry in infants should not be done at lower sampling flow rates.

Authors:  N Gravenstein
Journal:  J Clin Monit       Date:  1989-01

3.  Can we trust end-tidal carbon dioxide measurements in infants?

Authors:  F J Sasse
Journal:  J Clin Monit       Date:  1985-07

4.  Technical and clinical aspects of capnography in neonates.

Authors:  H Kirpalani; S Kechagias; J Lerman
Journal:  J Med Eng Technol       Date:  1991 Jul-Oct

5.  End-tidal PCO2 monitoring in infants and children ventilated with either a partial rebreathing or a non-rebreathing circuit.

Authors:  J M Badgwell; J E Heavner; W S May; J F Goldthorn; J Lerman
Journal:  Anesthesiology       Date:  1987-03       Impact factor: 7.892

6.  Accuracy of expiratory carbon dioxide measurements using the coaxial and circle breathing circuits in small subjects.

Authors:  R A Schieber; A Namnoum; A Sugden; A L Saville; R A Orr
Journal:  J Clin Monit       Date:  1985-07
  6 in total
  1 in total

1.  A novel application of capnography during controlled human exposure to air pollution.

Authors:  Karl Z Lukic; Bruce Urch; Michael Fila; Marie E Faughnan; Frances Silverman
Journal:  Biomed Eng Online       Date:  2006-10-18       Impact factor: 2.819

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.