Quantification of left ventricular volumes and ejection fraction using freehand transthoracic three-dimensional echocardiography: comparison with magnetic resonance imaging.

Herman F J Mannaerts1, Johannes A Van Der Heide, Otto Kamp, Theano Papavassiliu, Johannes T Marcus, Aernout Beek, Albert C Van Rossum, Jos Twisk, Cees A Visser.   

Abstract

OBJECTIVES: Our aim was to validate 3-dimensional echocardiography (3DE) for assessment of left ventricular (LV) end-diastolic volume, end-systolic volume (ESV), stroke volume, and ejection fraction (EF) using the freehand-acquisition method. Furthermore, LV volumes by breath hold-versus free breathing-3DE acquisition were assessed and compared with magnetic resonance imaging (MRI).
METHODS: From the apical position, a fan-like 3DE image was acquired during free breathing and another, thereafter, during breath hold. In 27 patients, 28 breath hold- and 24 free breathing-3DE images were acquired. A total of 17 patients underwent both MRI and 3DE. MRI contours were traced along the outer endocardial contour, including trabeculae, and along the inner endocardial contour, excluding trabeculae, from the LV volume.
RESULTS: All 28 (100%) breath hold- and 86% of free breathing-3DE acquisitions could be analyzed. Intraobserver variation (percentual bias +/- 2 SD) of end-diastolic volume, ESV, stroke volume, and EF for breath-hold 3DE was, respectively, 0.3 +/- 10.2%, 0.3 +/- 14.6%, 0.1 +/- 18.4%, and -0.1 +/- 5.8%. For free-breathing 3DE, findings were similar. A significantly better interobserver variability, however, was observed for breath-hold 3DE for ESV and EF. Comparison of breath-hold 3DE with MRI inner contour showed for end-diastolic volume, ESV, stroke volume, and EF, a percentual bias (+/- 2 SD) of, respectively, -13.5 +/- 26.9%, -17.7 +/- 47.8%, -10.6 +/- 43.6%, and -1.8 +/- 11.6%. Compared with the MRI outer contour, a significantly greater difference was observed, except for EF.
CONCLUSIONS: 3DE using the freehand method is fast and highly reproducible for (serial) LV volume and EF measurement, and, hence, ideally suited for clinical decision making and trials. Breath-hold 3DE is superior to free-breathing 3DE regarding image quality and reproducibility. Compared with MRI, 3DE underestimates LV volumes, but not EF, which is mainly explained by differences in endocardial contour tracing by MRI (outer contour) and 3DE (inner contour) of the trabecularized endocardium. Underestimation is reduced when breath-hold 3DE is compared with inner contour analysis of the MRI dataset.

Entities:  

Mesh:

Year:  2003        PMID: 12574735     DOI: 10.1067/mje.2003.7

Source DB:  PubMed          Journal:  J Am Soc Echocardiogr        ISSN: 0894-7317            Impact factor:   5.251


  20 in total

1.  Assessment of left ventricular volume and function by integration of simplified 3D echocardiography, tissue harmonic imaging and automated extraction of endocardial borders.

Authors:  Donato Mele; Roberta Teoli; Corrado Cittanti; Giovanni Pasanisi; Gabriele Guardigli; Robert A Levine; Roberto Ferrari
Journal:  Int J Cardiovasc Imaging       Date:  2004-06       Impact factor: 2.357

Review 2.  Novel techniques for assessment of left ventricular systolic function.

Authors:  Sonal Chandra; Hicham Skali; Ron Blankstein
Journal:  Heart Fail Rev       Date:  2011-07       Impact factor: 4.214

3.  Contrast-enhanced versus non-enhanced three-dimensional echocardiography of left ventricular volumes.

Authors:  J A van der Heide; H F J Mannaerts; L Yang; G T Sieswerda; C A Visser; O Kamp
Journal:  Neth Heart J       Date:  2008-02       Impact factor: 2.380

4.  Real-time three-dimensional echocardiographic assessment of left ventricular remodeling index in patients with hypertensive heart disease and coronary artery disease.

Authors:  Ming Chen; Jing Wang; Mingxing Xie; Xinfang Wang; Qing Lv; Lei Wang; Yan Li; Manli Fu
Journal:  J Huazhong Univ Sci Technolog Med Sci       Date:  2009-02-18

5.  Clinical application of three-dimensional echocardiography: past, present and future.

Authors:  S A Kleijn; O Kamp
Journal:  Neth Heart J       Date:  2009-01       Impact factor: 2.380

Review 6.  New advances in quantitative echocardiography.

Authors:  Steve L Liao; Mario J Garcia
Journal:  J Nucl Cardiol       Date:  2008 Mar-Apr       Impact factor: 5.952

7.  Accuracy of three-dimensional echocardiography in patients with prior anteroseptal myocardial infarction.

Authors:  Yoshiaki Maruyama; Kentaro Toyama; Shunichi Sato; Nobuyuki Masaki; Masato Kirimura; Nobuo Yoshimoto
Journal:  J Echocardiogr       Date:  2009-01-14

8.  Limitations in the current screening practice of assessing left ventricular ejection fraction for a primary prophylactic implantable defibrillator in southern Ontario.

Authors:  Christopher Lane; Paul Dorian; Nina Ghosh; Maria Radina; Suzan O'Donnell; Kevin Thorpe; Iqwal Mangat; Victoria Korley; Arnold Pinter
Journal:  Can J Cardiol       Date:  2010-03       Impact factor: 5.223

9.  Reproducibility of MRI measurements of right ventricular size and function in patients with normal and dilated ventricles.

Authors:  Christiaan F Mooij; Cornelis J de Wit; Dionne A Graham; Andrew J Powell; Tal Geva
Journal:  J Magn Reson Imaging       Date:  2008-07       Impact factor: 4.813

10.  Recommendations for cardiovascular magnetic resonance in adults with congenital heart disease from the respective working groups of the European Society of Cardiology.

Authors:  Philip J Kilner; Tal Geva; Harald Kaemmerer; Pedro T Trindade; Juerg Schwitter; Gary D Webb
Journal:  Eur Heart J       Date:  2010-01-11       Impact factor: 29.983

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.