Literature DB >> 12566645

Natural history and predictors of temporary epicardial pacemaker wire function in patients after open heart surgery.

Farhad Elmi1, Nicholas G Tullo, Koroush Khalighi.   

Abstract

STUDY
OBJECTIVES: To assess the short-term reliability of temporary epicardial pacing wires in patients after open heart surgery and to determine the influence of preoperative factors on the performance of these wires. PATIENTS AND METHODS: In this prospective, longitudinal cohort study, 60 patients (41 men) with a mean age of 66.8 +/- 8.9 years were studied for 5.3 +/- 1.1 days postoperatively. Each patient's history, physical findings, hemodynamics and echocardiographic data were analyzed. Postoperatively, capture threshold, the electrogram amplitude and slew rate, and the lead impedance were determined daily in both chambers.
RESULTS: Both atrial and ventricular thresholds increased significantly (p < 0.001) by the 4th postoperative day. The P-wave amplitudes were consistently less than half of the R-wave amplitudes. Both P- and R-wave amplitudes deteriorated significantly by the 2nd postoperative day (p < 0.001 and p < 0.02, respectively). Atrial and ventricular slew rates and lead impedance decreased significantly during the observation period (p < 0.002).
CONCLUSION: Although both atrial and ventricular temporary epicardial leads are reliable for short-term use, their function deteriorates on a daily basis. Perioperative factors are generally not helpful in predicting the performance of temporary epicardial pacing wires. Copyright 2002 S. Karger AG, Basel

Entities:  

Mesh:

Year:  2002        PMID: 12566645     DOI: 10.1159/000067312

Source DB:  PubMed          Journal:  Cardiology        ISSN: 0008-6312            Impact factor:   1.869


  3 in total

1.  Cardiac arrest caused by undersensing of a temporary epicardial pacemaker.

Authors:  D Chemello; A Subramanian; N Kumaraswamy
Journal:  Can J Cardiol       Date:  2010-01       Impact factor: 5.223

2.  Post operative temporary epicardial pacing: When, how and why?

Authors:  Anjan S Batra; Seshadri Balaji
Journal:  Ann Pediatr Cardiol       Date:  2008-07

3.  Predictors of temporary epicardial pacing wires use after valve surgery.

Authors:  Nizar R Alwaqfi; Khaled S Ibrahim; Yousef S Khader; Ahmad Abu Baker
Journal:  J Cardiothorac Surg       Date:  2014-02-12       Impact factor: 1.637