Determination of glucose in whole blood samples by mid-infrared spectroscopy.

Yoen-Joo Kim1, Sangjoon Hahn, Gilwon Yoon.   

Abstract

We have determined the glucose concentration of whole blood from mid-infrared spectra without sample preparation or use of chemical reagents. We selected 1119-1022 cm(-1) as the optimal wavelength range for our measurement by making a first-loading vector analysis based on partial least-squares regression. We examined the influence of hemoglobin on samples by using different calibration and prediction sets. The accuracy of glucose prediction depended on the hemoglobin level in the calibration model; the sample set should represent the entire range of hemoglobin concentration. We obtained an accuracy of 5.9% in glucose prediction, and this value is well within a clinically acceptable range.