Literature DB >> 12563140

MR image-guided endovascular procedures with the ultrasmall superparamagnetic iron oxide SH U 555 C as an intravascular contrast agent: study in pigs.

Frank K Wacker1, Klaus Reither, Wolfgang Ebert, Michael Wendt, Jonathan S Lewin, Karl-Juergen Wolf.   

Abstract

PURPOSE: To evaluate the feasibility of using the ultrasmall superparamagnetic iron oxide (USPIO) SH U 555 C as an intravascular contrast agent for magnetic resonance (MR) image-guided vascular procedures with an open MR imaging system.
MATERIALS AND METHODS: All experiments were performed with MR imaging at 0.2 T. MR image-guided interventions were performed in USPIO-enhanced vessels in four pigs. With near real-time MR image guidance (acquisition time, 0.64 second per section), the splenic and renal arteries were consecutively catheterized by using a susceptibility artifact-based catheter-guide wire combination. Angioplasty and stent implantation were performed four times in the renal artery and twice in the iliac artery. Intraaortal signal intensity (SI) was measured during the interventions.
RESULTS: After administration of SH U 555 C (40 micromol of iron per kilogram of body weight), a three-dimensional MR angiographic sequence was performed that allowed visualization of the abdominal and pelvic vessels that were as small as 2 mm in diameter. Catheterization, angioplasty, and stent implantation were successfully guided in the USPIO-enhanced vasculature. Sixty minutes after contrast agent injection, the mean aortic SI was 70% of the maximum measured enhancement levels.
CONCLUSION: One intravenous injection of SH U 555 C enabled long, continuous intravascular SI enhancement at MR angiography, and, in combination with susceptibility artifact-based device tracking, the injection allowed the performance of MR imaging-guided intravascular interventions in an open MR imaging system.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12563140     DOI: 10.1148/radiol.2262011815

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  3 in total

Review 1.  MR-guided endovascular interventions: a comprehensive review on techniques and applications.

Authors:  Sebastian Kos; Rolf Huegli; Georg M Bongartz; Augustinus L Jacob; Deniz Bilecen
Journal:  Eur Radiol       Date:  2007-12-11       Impact factor: 5.315

Review 2.  MR angiography with blood pool contrast agents.

Authors:  Jens Bremerich; Deniz Bilecen; Peter Reimer
Journal:  Eur Radiol       Date:  2007-07-17       Impact factor: 5.315

3.  Deuterium oxide as a contrast medium for real-time MRI-guided endovascular neurointervention.

Authors:  Lin Chen; Jing Liu; Chengyan Chu; Zheng Han; Nirhbay Yadav; Jiadi Xu; Renyuan Bai; Verena Staedtke; Monica Pearl; Piotr Walczak; Peter van Zijl; Miroslaw Janowski; Guanshu Liu
Journal:  Theranostics       Date:  2021-04-15       Impact factor: 11.556

  3 in total