The measurement and comparison of jet characteristics of surgical pulse lavage devices.

J Morgan1, G Holder, G Desoutter.   

Abstract

The characteristics of pulse waveform, pulse duration, pulse frequency, impact force/pressure, and flow rate of 4 commercially available pulse lavage devices are measured and compared. Validation of the measurements obtained is provided by a laboratory system that generates and measures precise jet waveforms. However, the value of the devices studied awaits clinical trials. Copyright 2003, Elsevier Science (USA). All rights reserved.