Stiffness characteristics and inter-fragmentary displacements with different hybrid external fixators.

Lang Yang1, S Nayagam, Michael Saleh.   

Abstract

OBJECTIVES: To determine how the different approaches of constructing a hybrid external skeletal fixator affect the mechanical environment at the fracture site.
DESIGN: A comparative biomechanical analysis of external fixators that represent a spectrum of designs from unilateral to circular fixators.
BACKGROUND: The most pertinent parameter that affects fracture healing is the inter-fragmentary displacement. Most studies on the mechanical performance of external fixators have concentrated on the overall stiffness of the fixators.
METHODS: Mechanical testing was performed on four types of hybrid fixators in which the overall stiffness of the fixator as well as the axial, transverse shear and angular displacements at the fracture site were measured.
RESULTS: The Ilizarov hybrid fixator with one wire and one screw on each ring behaved more like a unilateral fixator than a circular fixator. The bar-ring hybrid fixator that simply connected a unilateral fixator body to a wire/ring assembly was too flexible. Reinforcing the bar-ring hybrid system with diagonally placed struts eliminated the deformation associated with the unilateral body and the resulting construct performed considerably better. The two-ring hybrid fixator most closely resembled the Ilizarov fixator.
CONCLUSIONS: Depending on the construction characteristics hybrid fixators possess a spectrum of mechanical characteristics of both unilateral and circular fixators. The constructions using two rings in which at least one bone segment is supported entirely by wires appears to have the best mechanical characteristics. RELEVANCE: Understanding how the different approaches of constructing a hybrid fixator affect the mechanical environment at the fracture site will enable surgeons to chose or build the most appropriate fixator for each clinical situation.

Mesh:

Year:  2003        PMID: 12550816     DOI: 10.1016/s0268-0033(02)00175-4

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  15 in total

1.  The mechanics of external fixation.

Authors:  Austin T Fragomen; S Robert Rozbruch
Journal:  HSS J       Date:  2007-02

2.  What Are the Biomechanical Properties of the Taylor Spatial Frame™?

Authors:  Daniel J Henderson; Jeremy L Rushbrook; Paul J Harwood; Todd D Stewart
Journal:  Clin Orthop Relat Res       Date:  2016-11-28       Impact factor: 4.176

Review 3.  External fixators: looking beyond the hardware maze.

Authors:  Kimia Khalatbari Kani; Jack A Porrino; Felix S Chew
Journal:  Skeletal Radiol       Date:  2019-09-12       Impact factor: 2.199

Review 4.  The Ilizarov paradigm: thirty years with the Ilizarov method, current concerns and future research.

Authors:  Alexander V Gubin; Dmitry Y Borzunov; Tatiana A Malkova
Journal:  Int Orthop       Date:  2013-05-28       Impact factor: 3.075

5.  What Are the Biomechanical Effects of Half-pin and Fine-wire Configurations on Fracture Site Movement in Circular Frames?

Authors:  Daniel J Henderson; Jeremy L Rushbrook; Todd D Stewart; Paul J Harwood
Journal:  Clin Orthop Relat Res       Date:  2015-12-07       Impact factor: 4.176

6.  Biomechanical Analysis of the Behaviour at the Metaphyseal-Diaphyseal Junction of Complex Tibial Plateau Fractures Using Two Circular Fixator Configurations.

Authors:  Gracielle S Cardoso; Renato Amorim; Francisco M Penha; Françoá J Horn; Carlos Rm Roesler; Jefferson Lb Marques
Journal:  Strategies Trauma Limb Reconstr       Date:  2020 Sep-Dec

7.  Direct and indirect loading of the Ilizarov external fixator: the effect on the interfragmentary movements and compressive loads.

Authors:  Jan Gessmann; Hinnerk Baecker; Birger Jettkant; Gert Muhr; Dominik Seybold
Journal:  Strategies Trauma Limb Reconstr       Date:  2011-02-11

8.  Comparison of transverse wires and half pins in Taylor Spatial Frame: a biomechanical study.

Authors:  Ashish Khurana; Carlton Byrne; Sam Evans; Hiro Tanaka; Kartik Haraharan
Journal:  J Orthop Surg Res       Date:  2010-03-27       Impact factor: 2.359

9.  The influence of a weight-bearing platform on the mechanical behavior of two Ilizarov ring fixators: tensioned wires vs. half-pins.

Authors:  Jan Gessmann; Mustafa Citak; Birger Jettkant; Thomas A Schildhauer; Dominik Seybold
Journal:  J Orthop Surg Res       Date:  2011-12-12       Impact factor: 2.359

10.  Hybrid external fixation for neglected fractures of the distal radius: results after one year.

Authors:  Paweł Grala; Wojciech Zieliński
Journal:  J Orthop Traumatol       Date:  2008-09-19
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.