Literature DB >> 12548459

[How do metallic middle ear implants behave in the MRI?].

P Kwok1, A Waldeck, J Strutz.   

Abstract

BACKGROUND: Magnetic resonance imaging (MRI) has gained in frequency and importance as a diagnostic procedure. In respect to the close anatomical relationship in the temporal bone it is necessary to know whether it is hazardous to patients with metallic middle ear implants regarding displacement and rise in temperature. For the MR image quality artefacts caused by metallic prostheses should be low.
METHODS: Four different stapes prostheses made from titanium, gold, teflon/platinum and teflon/steel, a titanium total ossicular reconstruction prosthesis (TORP) and two ventilation tubes (made from titanium and gold) were tested in a 1.5 Tesla MRI machine regarding their displacement. All objects were first placed in a petri dish, then suspended from a thread and finally immersed in a dish filled with Gadolinium. Temperature changes of the implants were recorded by a pyrometer.
RESULTS: None of the implants moved when they were placed in the petri dish or suspended from the thread. On the water surface the teflon/platinum and the teflon/steel pistons adjusted their direction with their axis longitudinally to the MRI scanner opening and the teflon/steel piston floated towards the MRI-machine when put close enough to the scanner opening. No rise in temperature was recorded. All implants showed as little artefacts that would still make an evaluation of the surrounding tissue possible.
CONCLUSION: Patients with any of the metallic middle ear implants that were examined in this study may undergo MRI-investigations without significant adverse effects.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12548459     DOI: 10.1055/s-2003-36905

Source DB:  PubMed          Journal:  Laryngorhinootologie        ISSN: 0935-8943            Impact factor:   1.057


  5 in total

1.  [Magnetic resonance angiography of the head and neck area].

Authors:  R Aschenbach; D Esser
Journal:  HNO       Date:  2004-01       Impact factor: 1.284

2.  [Effects of magnetic resonance tomography on upper eyelid implants].

Authors:  T Schrom; H Bauknecht; A Berghaus; H Scherer
Journal:  HNO       Date:  2005-08       Impact factor: 1.284

3.  Behavior of metal implants used in ENT surgery in 7 Tesla magnetic resonance imaging.

Authors:  Ariane Thelen; Hans-Christian Bauknecht; Patrick Asbach; Thomas Schrom
Journal:  Eur Arch Otorhinolaryngol       Date:  2006-07-12       Impact factor: 2.503

4.  Passive and active middle ear implants.

Authors:  Dirk Beutner; Karl-Bernd Hüttenbrink
Journal:  GMS Curr Top Otorhinolaryngol Head Neck Surg       Date:  2011-03-10

5.  Anatomical and functional results of ossiculoplasty using titanium prosthesis.

Authors:  G Lahlou; G Sonji; D De Seta; I Mosnier; F Y Russo; O Sterkers; D Bernardeschi
Journal:  Acta Otorhinolaryngol Ital       Date:  2018-08       Impact factor: 2.124