Literature DB >> 12544771

Whole blood international normalization ratio measurements in children using near-patient monitors.

William L Nowatzke1, Michael Landt, Carl Smith, Timothy Wilhite, Charles Canter, Lori Luchtman-Jones.   

Abstract

PURPOSE: To report a comparison of international normalization ratio (INR) measurements on four near-patient (point-of-care or bedside) whole blood INR monitors in children. PATIENTS AND METHODS: The INR results from 19 ambulatory pediatric subjects (30 hospital visits) receiving warfarin sodium were analyzed on four near-patient monitors and compared with plasma INR measurements on the laboratory CA-1000 Analyze. The instruments evaluated were CoaguChek, Hemochron Jr. Signature, ProTime Microcoagulation System, and RapidpointCoag.
RESULTS: The INR measurements ranged from 1.05 to 5.25. Over the entire INR range, the near-patient instrument with the least bias relative to the CA-1000 was the RapidpointCoag (r(2) = 0.923). The correlations (r(2)) of the CoaguCheck, Hemochron Jr., and ProTime were 0.877, 0.834, and 0.885, respectively. Precision studies involved repeated analysis of one nonmedicated adult (mean CA-1000 INR = 0.908) and one adult receiving oral anticoagulation therapy (mean CA-1000 INR = 2.42). The coefficient of variation on the near-patient monitors for both adult volunteers ranged from 4.9% to 22.3%. Bilirubin levels up to 20 mg/dL did not interfere in any of the methods.
CONCLUSIONS: Near-patient testing whole blood INR monitors offer acceptably accurate and precise measurements. Values obtained on near-patient monitors may vary considerably from the reference method, and data obtained should serve as a supplement to, but not a replacement for, routine clinical laboratory measurements.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12544771     DOI: 10.1097/00043426-200301000-00008

Source DB:  PubMed          Journal:  J Pediatr Hematol Oncol        ISSN: 1077-4114            Impact factor:   1.289


  4 in total

1.  On-chip titration of an anticoagulant argatroban and determination of the clotting time within whole blood or plasma using a plug-based microfluidic system.

Authors:  Helen Song; Hung-Wing Li; Matthew S Munson; Thuong G Van Ha; Rustem F Ismagilov
Journal:  Anal Chem       Date:  2006-07-15       Impact factor: 6.986

2.  Validity of CoaguChek S for home monitoring of anticoagulant therapy in pediatrics.

Authors:  James Hill; Sébastien Perreault; Michel Dorval
Journal:  Can J Cardiol       Date:  2007-01       Impact factor: 5.223

3.  Antithrombotic therapy in neonates and children: Antithrombotic Therapy and Prevention of Thrombosis, 9th ed: American College of Chest Physicians Evidence-Based Clinical Practice Guidelines.

Authors:  Paul Monagle; Anthony K C Chan; Neil A Goldenberg; Rebecca N Ichord; Janna M Journeycake; Ulrike Nowak-Göttl; Sara K Vesely
Journal:  Chest       Date:  2012-02       Impact factor: 9.410

4.  International normalized ratio testing with point-of-care coagulometer in healthy term neonates.

Authors:  Shigeo Iijima; Toru Baba; Daizo Ueno; Akira Ohishi
Journal:  BMC Pediatr       Date:  2014-07-09       Impact factor: 2.125

  4 in total