Literature DB >> 12541257

Theoretical and experimental evaluation of detached endcaps for 3 T birdcage coils.

Marcello Alecci1, Christopher M Collins, James Wilson, Wanzhan Liu, Michael B Smith, Peter Jezzard.   

Abstract

The use of detached endcaps for 3 T birdcage coils was investigated both theoretically and experimentally. Finite difference time domain analysis, along with workbench and MRI techniques, were used to map the radiofrequency (RF) B(1) distribution along the coil axis with and without an endcap. Without an endcap the measured B(1) value at the service end of the birdcage was only 45% of the value at the coil's center. This was improved to 85% with a detached endcap of maximum achievable diameter (375 mm), positioned 4 mm from the RF shield. The B(1) field distribution on the patient side of the coil was unaffected by the presence of the endcap. The dependence of the B(1) distribution as a function of endcap diameter was also investigated. Surprisingly, simulations and experiments show that there is an optimum ratio of endcap-to-birdcage coil diameter (approximately 1.08) that gives the best B(1) homogeneity. In the human head the optimized endcap, positioned 16 mm from the RF shield, improves the MRI signal amplitude from 55% to 85% of maximum toward the service end. This novel endcap design is easy to implement with existing birdcage coils, and could prove useful when flexibility in access to the RF coil is required. Copyright 2003 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2003        PMID: 12541257     DOI: 10.1002/mrm.10352

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  3 in total

1.  Sodium MR imaging detection of mild Alzheimer disease: preliminary study.

Authors:  E A Mellon; D T Pilkinton; C M Clark; M A Elliott; W R Witschey; A Borthakur; R Reddy
Journal:  AJNR Am J Neuroradiol       Date:  2009-02-12       Impact factor: 3.825

Review 2.  Numerical field calculations considering the human subject for engineering and safety assurance in MRI.

Authors:  Christopher M Collins
Journal:  NMR Biomed       Date:  2009-11       Impact factor: 4.044

3.  An Asymmetric Birdcage Coil for Small-animal MR Imaging at 7T.

Authors:  Kyoung-Nam Kim; Sang-Doc Han; Jeung-Hoon Seo; Phil Heo; Dongkyeom Yoo; Geun Ho Im; Jung Hee Lee
Journal:  Magn Reson Med Sci       Date:  2016-09-30       Impact factor: 2.471

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.