Thomas K Aldrich1, Mark Moosikasuwan, Samir D Shah, Kedar S Deshpande.
Abstract
To determine whether blood hemoglobin concentration ([Hb]) could be measured noninvasively as the ratio of pulsatile changes in attenuation (absorbance plus scatter) of light (D) across a body part to changes in light path length (l), we measured transmission of near-infrared light (905 +/- 10 nm) through a finger, using a modified pulse oximeter, and simultaneously monitored fingertip diameter, using a sonomicrometer. In 25 subjects with [Hb] ranging from 3.1 to 18.2 gm/dl, and with normal oxygenation, average D/l ratio over 30-60 s correlated strongly with [Hb] measured by Coulter counter (r = 0.84, p << 0.001), though with considerable scatter, with absolute value of differences averaging 17% of the mean. Using 12 gm/dl and 0.75 mm(-1) as the lower limits of normal for [Hb] and D/l, respectively, two of nine normals had low (D/l) (78% specificity), and only one of 16 anemic subjects had borderline normal (D/l) (94%-100% sensitivity). The positive predictive value of a low (D/l) was 88% and the negative predictive value was 87.5%. With further development, this technique may reduce the need for phlebotomy, thereby reducing risks and costs and improving the experience of being a patient.Entities:
Mesh:
Substances:
Year: 2002 PMID: 12540205 DOI: 10.1114/1.1527046
Source DB: PubMed Journal: Ann Biomed Eng ISSN: 0090-6964 Impact factor: 3.934