Literature DB >> 12534714

Blood compatible design of a pulsatile blood pump using computational fluid dynamics and computer-aided design and manufacturing technology.

Eiji Okamoto1, Takuya Hashimoto, Taku Inoue, Yoshinori Mitamura.   

Abstract

Thrombus formation is a critical issue when designing a long-term implantable left ventricular assist system (LVAS). Fluid dynamic characteristics of blood flow are one of the main factors that cause thrombus formation. In this study, we optimized the fluid dynamics of a sac blood pump in our LVAS to ensure minimization of shear-related blood damage that could lead to thrombus formation. A pump housing and a sac chamber were designed with computer-aided design (CAD) software, and fluid dynamics were estimated by computational fluid dynamic (CFD) analysis. We adopted distribution of CFD results for qualitative evaluation, and we also tried to estimate normalized index of hemolysis (NIH) from the results of CFD analysis as a quantitative index of optimization for geometry of the blood pump chamber. A prototype model of the optimized blood pump was made using a three-axis computer machine tool by whittling pieces of nonfoamed polyurethane. Shear stress and theoretical NIH in the redesigned model were lower than those in the first model. Area of flow stagnation that was observed in the first model was not seen in the redesigned model. The results demonstrate that application of CAD/CAM technology to design an artificial heart contributes to optimizing a blood pump chamber for the purpose of reducing thrombus formation.

Entities:  

Mesh:

Year:  2003        PMID: 12534714     DOI: 10.1046/j.1525-1594.2003.07183.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  3 in total

Review 1.  The use of computational fluid dynamics in the development of ventricular assist devices.

Authors:  Katharine H Fraser; M Ertan Taskin; Bartley P Griffith; Zhongjun J Wu
Journal:  Med Eng Phys       Date:  2010-11-13       Impact factor: 2.242

2.  A Cellular Model of Shear-Induced Hemolysis.

Authors:  Salman Sohrabi; Yaling Liu
Journal:  Artif Organs       Date:  2017-01-03       Impact factor: 3.094

3.  Coarse-Grained Modeling of Pore Dynamics on the Red Blood Cell Membrane under Large Deformations.

Authors:  Meghdad Razizadeh; Mehdi Nikfar; Ratul Paul; Yaling Liu
Journal:  Biophys J       Date:  2020-06-24       Impact factor: 4.033