Literature DB >> 12532244

Assessment of the DIXTAL DX-2710 automated oscillometric device for blood pressure measurement with the validation protocols of the British Hypertension Society (BHS) and the Association for the Advancement of Medical Instrumentation (AAMI).

Gisele M P Mano1, Viviane F Souza, Angela M G Pierin, Josiane C Lima, Edna C Ignes, Kátia C Ortega, Décio Mion.   

Abstract

OBJECTIVE: To assess the Dixtal DX2710 automated oscillometric device used for blood pressure measurement according to the protocols of the BHS and the AAMI.
METHODS: Three blood pressure measurements were taken in 94 patients (53 females 15 to 80 years). The measurements were taken randomly by 2 observers trained to measure blood pressure with a mercury column device connected with an automated device. The device was classified according to the protocols of the BHS and AAMI. RESULT: The mean of blood pressure levels obtained by the observers was 148+/-38/93+/-25 mmHg and that obtained with the device was 148+/-37/89+/-26 mmHg. Considering the differences between the measurements obtained by the observer and those obtained with the automated device according to the criteria of the BHS, the following classification was adopted: "A" for systolic pressure (69% of the differences < 5; 90% < 10; and 97% < 15 mmHg); and "B" for diastolic pressure (63% of the differences < 5; 83% < 10; and 93% < 15 mmHg). The mean and standard deviation of the differences were 0+/-6.27 mmHg for systolic pressure and 3.82+/-6.21 mmHg for diastolic pressure.
CONCLUSION: The Dixtal DX2710 device was approved according to the international recommendations.

Entities:  

Mesh:

Year:  2002        PMID: 12532244     DOI: 10.1590/s0066-782x2002001500006

Source DB:  PubMed          Journal:  Arq Bras Cardiol        ISSN: 0066-782X            Impact factor:   2.000


  5 in total

1.  Plasma renin and cardiovascular responses to the cold pressor test differ in black and white populations: The SABPA study.

Authors:  L F Gafane; R Schutte; J M Van Rooyen; A E Schutte
Journal:  J Hum Hypertens       Date:  2015-08-27       Impact factor: 3.012

2.  Peripheral Vascular Resistance Impairment during Isometric Physical Exercise in Normotensive Offspring of Hypertensive Parents.

Authors:  Natália Portela; Josária Ferraz Amaral; Pedro Augusto de Carvalho Mira; Livia Victorino de Souza; Daniel Godoy Martinez; Mateus Camaroti Laterza
Journal:  Arq Bras Cardiol       Date:  2017-07-10       Impact factor: 2.000

3.  How to avoid discontinuation of antihypertensive treatment: The experience in São Paulo, Brazil.

Authors:  Katia Coelho Ortega; Josiane Lima de Gusmão; Angela Maria Geraldo Pierin; José Luiz Nishiura; Edna Caetano Ignez; Carlos Alexandre Segre; Carlucci Gualberto Ventura; Gisele Peixoto Mano; Viviane Fontes; Francisco Mogadouro da Cunha; Décio Mion
Journal:  Clinics (Sao Paulo)       Date:  2010       Impact factor: 2.365

4.  Oral Fludrocortisone Test for Salt-Sensitive Screening in Hypertensive Patients: A Randomized Crossover Trial.

Authors:  Andrea Pio-Abreu; Isac de Castro; Giovanio Vieira da Silva; Katia Coelho Ortega; Decio Mion
Journal:  Int J Hypertens       Date:  2018-11-15       Impact factor: 2.420

5.  Health-related quality of life and blood pressure control in hypertensive patients with and without complications.

Authors:  Josiane Lima de Gusmão; Decio Mion; Angela Maria Geraldo Pierin
Journal:  Clinics (Sao Paulo)       Date:  2009       Impact factor: 2.365

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.