Literature DB >> 12531676

A comparison of ex vivo and in vitro Sutter metacarpophalangeal prostheses.

T J Joyce1, R H Milner, A Unsworth.   

Abstract

Forty-one Sutter metacarpophalangeal prostheses were implanted into 11 hands of nine patients. Twelve of these prostheses were revised from three patients after a mean period of 42 months. Of the 12 prostheses, 11 showed fracture at the junction of the distal stem and the central hinge region. Two Sutter metacarpophalangeal prostheses were tested on a single station finger simulator and both failed due to fracture at the junction of the hinge and the distal stem.

Entities:  

Mesh:

Year:  2003        PMID: 12531676     DOI: 10.1054/jhsb.2002.0862

Source DB:  PubMed          Journal:  J Hand Surg Br        ISSN: 0266-7681


  2 in total

1.  Wear testing of a DJOA finger prosthesis in vitro.

Authors:  Thomas J Joyce
Journal:  J Mater Sci Mater Med       Date:  2010-02-11       Impact factor: 3.896

2.  Pyrocarbon finger joint implant: an evidence-based analysis.

Authors: 
Journal:  Ont Health Technol Assess Ser       Date:  2004-03-01
  2 in total