Abstract
Several probes for measuring dissolved carbon dioxide (CO2) concentration were installed in a 68-litre fermentor and their effectiveness compared. Submerged silastic rubber tubing gave reproducible results over a wide range of operating conditions and was generally superior to all other probes evaluated. The silastic rubber probe was used to compare the partial pressure of CO2 in viscous fermentation media with that in the fermentor exhaust gas. No significant difference was found. Results show that determination of the CO2 partial pressure in the exhaust gas gives an excellent approximation of the partial pressure of dissolved CO2 in the liquid medium, eliminating the need for measurement of CO2 concentration in the broth.Entities:
Mesh:
Substances:
Year: 1976 PMID: 1252990 DOI: 10.1139/m76-007
Source DB: PubMed Journal: Can J Microbiol ISSN: 0008-4166 Impact factor: 2.419