Literature DB >> 12527246

Cyclic joint loading can affect the initial stability of meniscal fixation implants.

Lutz Dürselen1, Jan Schneider, Mario Galler, Lutz E Claes, Gerhard Bauer.   

Abstract

OBJECTIVE: To identify the effect of cyclic joint loads on the initial fixation strength of different meniscal repair techniques.
DESIGN: Three different meniscal refixation implants and one suturing technique were tested for initial stability in porcine knee joints after cyclic joint loading.
BACKGROUND: Existing studies tested the initial pull out strength of meniscal fixation implants. It remains unknown how such implants perform under cyclic joint loading.
METHODS: The Meniscal Fastener, the Meniscal Arrow, the ClearFix Screw, and a horizontal suture (PDS 1) were implanted in 64 porcine knee joints after setting a tear in the posterior horn of the medial meniscus. In one group of each device the menisci were excised directly after implantation and subsequently subjected to a pull out test. The knee joints of the other groups were exposed to 5000 axial load cycles between 10 and 200 N prior to the pull out test.
RESULTS: The horizontal suture showed the highest pull out force (103 N, SD 19 N). Arrow (52 N, SD 18 N), Meniscal Fastener (29 N, SD 3 N), and Meniscal Screw (22 N, SD 8 N) failed at significantly lower loads. Cyclic loading led to a decreased initial pull out force only for the horizontal suture (82 N, SD 26 N) and the Meniscal Fastener (23 N, SD 5N).
CONCLUSIONS: Cyclic joint loading can lead to reduced fixation strength especially of meniscal refixation implants and sutures with relatively low stiffness. However, it remains unclear which fixation strength is necessary to provide conditions for meniscal healing. RELEVANCE: The current study adds information to the understanding of how meniscal fixation systems perform mechanically especially after cyclic joint loads.

Entities:  

Mesh:

Year:  2003        PMID: 12527246     DOI: 10.1016/s0268-0033(02)00139-0

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  4 in total

1.  A new rigid biodegradable anchor for meniscus refixation: biomechanical evaluation.

Authors:  Thore Zantop; Anne Kathleen Eggers; Volker Musahl; Andre Weimann; Joachim Hassenpflug; Wolf Petersen
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2003-12-19       Impact factor: 4.342

2.  The concept of three-dimensional hold of both circumferential and radial collagen fibres of the meniscus.

Authors:  Ashraf Abdelkafy
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2013-02-12       Impact factor: 4.342

Review 3.  Meta-analysis on biomechanical properties of meniscus repairs: are devices better than sutures?

Authors:  Daniel M Buckland; Patrick Sadoghi; Matthias D Wimmer; Patrick Vavken; Geert I Pagenstert; Victor Valderrabano; Claudio Rosso
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-05-20       Impact factor: 4.342

4.  Short- to mid-term results of arthroscopic meniscal repair of long vertical longitudinal tears using combined cruciate and horizontal suture techniques: a retrospective study.

Authors:  Ashraf Abdelkafy
Journal:  Eur J Orthop Surg Traumatol       Date:  2014-05-28
  4 in total