Mechanical failure of prosthetic human implants: a 10-year experience with aortic stent graft devices.

Tikva S Jacobs1, Jamie Won, Edwin C Gravereaux, Peter L Faries, Nicholas Morrissey, Victoria J Teodorescu, Larry H Hollier, Michael L Marin.   

Abstract

OBJECTIVE: The first endovascular stent graft was implanted to treat an abdominal aortic aneurysm more than a decade ago. This technique has evolved dramatically with the growing understanding of metallurgic and fabric sciences and improved device designs. However the potential for stent graft material failure remains. This investigation describes the incidence of material failure, potential modes of device fatigue, and the clinical significance of these failures.
METHODS: Six hundred eighty-six endovascular stent grafts were used to treat patients with aortic aneurysms. Device fatigue in the form of stent, suture fracture, or graft wear was identified with an analysis of follow-up radiographs and explanted stent grafts. A review of patient clinical histories, spiral computed tomographic scan studies, scanning electron microscopy, and energy dispersion spectroscopy of explanted devices was conducted to evaluate the modes and consequences of failure.
RESULTS: Sixty patients were identified with device fatigue, 49 of whom had abdominal endovascular repairs and 11 of whom had thoracic repairs. Of the 60 patients with stent graft fatigue, 43 patients had metallic stent fractures, 14 had suture disruptions, and three had graft holes. These material failures occurred within seven distinct stent graft designs. The average time to the recognition of failure was 19 months, with a mean follow-up period of 8 months since the event was identified. Eleven patients died, and one was lost to follow-up 2 years after identification of a stent fracture. The remaining patients are presently being followed eoyj physical examination, plain film radiograph, and computed tomographic scans for clinical sequelae of device fatigue.
CONCLUSION: Endovascular stent graft fatigue has been recognized in numerous devices after aortic implantation. Fatigue may take the form of stent, graft, or suture failure, with certain modes unique to specific stent graft devices. The clinical significance of stent graft material failure remains uncertain.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12514573     DOI: 10.1067/mva.2003.58

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  10 in total

1.  Successful endovascular management of an aortic rupture following stent placement for severe atherosclerotic stenosis: A case report.

Authors:  Thomas L Chung; Dipankar Mukherjee
Journal:  Int J Angiol       Date:  2007

2.  In vivo deformation of the human abdominal aorta and common iliac arteries with hip and knee flexion: implications for the design of stent-grafts.

Authors:  Gilwoo Choi; Lewis K Shin; Charles A Taylor; Christopher P Cheng
Journal:  J Endovasc Ther       Date:  2009-10       Impact factor: 3.487

3.  Thoracic endovascular aortic repair with homemade stent grafts: early and mid-term outcomes.

Authors:  Seiji Onitsuka; Atsuhisa Tanaka; Hidetoshi Akashi; Keiichi Akaiwa; Keita Mikasa; Kurando Kanaya; Yusuke Shintani; Shinichi Hiromatsu; Shigeaki Aoyagi
Journal:  Surg Today       Date:  2011-07-20       Impact factor: 2.549

Review 4.  Intravascular foreign bodies: danger of unretrieved fragmented medical devices.

Authors:  Minori Tateishi; Yasuko Tomizawa
Journal:  J Artif Organs       Date:  2009-06-18       Impact factor: 1.731

5.  Endovascular stent graft repair of abdominal and thoracic aortic aneurysms: a ten-year experience with 817 patients.

Authors:  Michael L Marin; Larry H Hollier; Sharif H Ellozy; David Spielvogel; Harold Mitty; Randall Griepp; Robert A Lookstein; Alfio Carroccio; Nicholas J Morrissey; Victoria J Teodorescu; Tikva S Jacobs; Michael E Minor; Claudie M Sheahan; Kristina Chae; Juliana Oak; Andrew Cha
Journal:  Ann Surg       Date:  2003-10       Impact factor: 12.969

6.  Finite volume analysis of temperature effects induced by active MRI implants: 2. Defects on active MRI implants causing hot spots.

Authors:  Martin H J Busch; Wolfgang Vollmann; Dietrich H W Grönemeyer
Journal:  Biomed Eng Online       Date:  2006-05-26       Impact factor: 2.819

Review 7.  Endovascular stent-graft placement for vascular failure of the thoracic aorta.

Authors:  Yoshihiko Kurimoto; Kiyofumi Morishita; Yasufumi Asai
Journal:  Vasc Health Risk Manag       Date:  2006

8.  Mid-Term Outcomes of Stent Overlap in Long Total Occluded Lesions of Superficial Femoral Artery.

Authors:  Ulas Bildirici; Mujdat Aktas; Emir Dervis; Umut Celikyurt
Journal:  Med Sci Monit       Date:  2017-06-26

9.  Rhombic-Shaped Channel Stent with Enhanced Drug Capacity and Fatigue Life.

Authors:  Hao-Ming Hsiao; Cheng-Han Lin; Yung-Kang Shen; Tzu-Yun Chou; Yen-Yu Hsu
Journal:  Micromachines (Basel)       Date:  2017-12-24       Impact factor: 2.891

10.  Research and clinical translation of trilayer stent-graft of expanded polytetrafluoroethylene for interventional treatment of aortic dissection.

Authors:  Gang Wang; Caiyun Gao; Benhao Xiao; Jie Zhang; Xunyuan Jiang; Qunsong Wang; Jingzhen Guo; Deyuan Zhang; Jianxiong Liu; Yuehui Xie; Chang Shu; Jiandong Ding
Journal:  Regen Biomater       Date:  2022-07-22
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.