The commissioning and quality assurance of the Automatic Positioning System on the Leksell gamma knife.

Tracy Soanes1, Anna Hampshire, Paul Vaughan, Christopher Brownett, Jeremy Rowe, Matthias Radatz, Andras Kemeny, Lee Walton.   

Abstract

OBJECT: The authors describe a series of commissioning checks that were developed following the installation of the Automatic Positioning System (APS) on the model C gamma knife.
METHODS: System reliability was checked by performing a series of test treatments using the APS. A phantom was designed to enable the exposure of small pieces of Gaf Chromic film at 40 different predefined x, y, and z coordinates. The phantom consisted of a base plate with a series of film holders to facilitate the exposures using a 4-mm field, the center of which was marked. A spreadsheet calculation was performed to verify the conversion from Leksell coordinates to the APS coordinates when the treatment angle (gamma angle) is other than 90 degrees. A number of APS plans were prepared and the coordinate transformation verified. Precision measurements were performed to verify the correct positioning of the high bars when attached to the frame. The Gaf Chromic films were exposed, and the APS plans were used when confined to positions within the high bar range. A test tool to verify accurate location of the high bars on the frame was also designed.
CONCLUSIONS: The performance of APS was verified independently of the manufacturer by using specially designed tools, phantoms, and spread sheets. At all points tested, the positional accuracy was found to be within specification. Conversion to APS coordinates was verified as correct.

Entities:  

Mesh:

Year:  2002        PMID: 12507099     DOI: 10.3171/jns.2002.97.supplement

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  2 in total

1.  American College of Radiology (ACR) and American Society for Radiation Oncology (ASTRO) Practice Guideline for the Performance of Stereotactic Radiosurgery (SRS).

Authors:  Steven K Seung; David A Larson; James M Galvin; Minesh P Mehta; Louis Potters; Christopher J Schultz; Santosh V Yajnik; Alan C Hartford; Seth A Rosenthal
Journal:  Am J Clin Oncol       Date:  2013-06       Impact factor: 2.339

2.  Characterization and commissioning of a Leksell Gamma Knife ICON system for framed and frameless stereotactic radiosurgery.

Authors:  Yue-Houng Hu; Susannah V Hickling; Jing Qian; C Robert Blackwell; Luke B McLemore; Erik J Tryggestad
Journal:  J Appl Clin Med Phys       Date:  2022-01-22       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.