A new magnetic resonance head coil and head immobilization device for gamma knife radiosurgery: an analysis of geometric distortion and signal/noise characteristics.

Paul Jursinic1, Robert Prost, Christopher Schultz.   

Abstract

OBJECT: The authors report on a new head coil into which the Leksell aluminum localization frame can be easily and securely mounted. Mechanically, the head coil interferes little with the patient.
METHODS: The head coil, which is for magnetic resonance (MR) imaging, is a 12-element quadrature transmit-and-receive high-pass birdcage coil with a nominal operation frequency (63.86 MHz). The coil was built into a plastic housing. This new head coil minimizes patient motion and provides a 20% increase in signal/noise ratios compared with standard head coils. An MR image test phantom was mounted in the coil and this allowed quantification of image distortion due to inhomogeneities in the main magnetic field, nonlinearity in the gradient field, and paramagnetism of the aluminum headframe. There were no significant differences in geometric distortion between the new head coil and the standard coil.
CONCLUSIONS: The new head coil has advantages for reducing patient movement artifacts and has a better signal/noise ratio with no reduction in geometric accuracy.

Entities:  

Mesh:

Year:  2002        PMID: 12507097     DOI: 10.3171/jns.2002.97.supplement

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  1 in total

1.  Geometric accuracy of 3D coordinates of the Leksell stereotactic skull frame in 1.5 Tesla- and 3.0 Tesla-magnetic resonance imaging: a comparison of three different fixation screw materials.

Authors:  Hisato Nakazawa; Yoshimasa Mori; Osamu Yamamuro; Masataka Komori; Yuta Shibamoto; Yukio Uchiyama; Takahiko Tsugawa; Masahiro Hagiwara
Journal:  J Radiat Res       Date:  2014-07-17       Impact factor: 2.724

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.