Retrieval of lost microcalcifications during stereotactic vacuum-assisted core biopsy.

Paul D Friedman1, Linda M Sanders, Christine Menendez, Lester Kalisher, Gina Petrillo.   

Abstract

OBJECTIVE: The purpose of our report was to describe patients in whom calcifications in the breast that were unequivocally removed during stereotactic core biopsy using the Mammotome device were not detected on the initial specimen radiograph. The lost calcifications in each instance were subsequently found when the tubing and contents of the debris canister were strained through a nonadhering dressing and radiographed. Additional situations in which calcifications are not seen on the initial specimen radiograph are described and recommendations are made.
CONCLUSION: When vacuum-assisted core biopsy procedures are performed, it is important to be aware of the possibility that calcifications may be aspirated into the debris canister, thus compromising the accuracy of the histopathologic diagnosis. We recommend changing the tubing and the debris canister after each procedure and, in certain situations, sending the strained canister contents to pathology for evaluation.

Entities:  

Mesh:

Year:  2003        PMID: 12490519     DOI: 10.2214/ajr.180.1.1800275

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  3 in total

1.  Minimal invasive surgery for gynecomastia - A novel approach.

Authors:  Oc Iwuagwu; Pj Drew
Journal:  Can J Plast Surg       Date:  2004

2.  Values of pathological analysis of lost tissue fragments in the vacuum canister during a vacuum-assisted stereotactic biopsy of the breast.

Authors:  M El Khoury; B Mesurolle; A Omeroglu; A Aldis; E Kao
Journal:  Br J Radiol       Date:  2013-03-21       Impact factor: 3.039

3.  In patients with DCIS: is it sufficient to histologically examine only those tissue specimens that contain microcalcifications?

Authors:  Alexander Poellinger; Susanne Diekmann; Ekkehart Dietz; Ulrich Bick; Felix Diekmann
Journal:  Eur Radiol       Date:  2008-01-08       Impact factor: 5.315

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.