Inflationary oscillometric blood pressure monitoring: validation of the OMRON-MIT.

Moneli Golara1, Clare Jones, Manjit Randhawa, Andrew H Shennan.   

Abstract

BACKGROUND: To evaluate the accuracy of the OMRON-MIT, (an automated inflationary oscillometric blood pressure monitoring device) for use in the general population by direct comparison to standard auscultatory mercury sphygmomanometry.
DESIGN: Two trained, blinded observers taking simultaneous blood pressure measurements alternating between a mercury sphygmomanometer and the OMRON-MIT.
METHODS: The evaluation was carried out according to the revised British Hypertension Society protocol. The validation data was also classified by the American Association for the Advancement of Medical Instrumentation (AAMI).
RESULTS: The OMRON-MIT device achieved a BHS grade A for both systolic blood pressure (SBP) and diastolic blood pressure (DBP). It also met the criteria for the AAMI protocol, the mean difference between the MIT and standard was -2.0 +/- 7 mmHg for SBP and -2.0 +/- 6 mmHg for DBP.
CONCLUSIONS: Inflationary oscillometry is a valid technique for the measurement of blood pressure. The OMRON-MIT, which uses this technique, is accurate and can be recommended for clinical use in the adult population. Copyright 2002 Lippincott Williams & Wilkins

Entities:  

Mesh:

Year:  2002        PMID: 12488653     DOI: 10.1097/00126097-200212000-00006

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  3 in total

1.  Validation of inflationary non-invasive blood pressure monitoring in adult surgical patients.

Authors:  Jun Onodera; Yoshifumi Kotake; Mitsue Fukuda; Rie Yasumura; Fujiko Oda; Nobukazu Sato; Ryoichi Ochiai; Takashi Usuda; Naoki Kobayashi; Sunao Takeda
Journal:  J Anesth       Date:  2010-12-28       Impact factor: 2.078

Review 2.  Sources of inaccuracy in the measurement of adult patients' resting blood pressure in clinical settings: a systematic review.

Authors:  Noa Kallioinen; Andrew Hill; Mark S Horswill; Helen E Ward; Marcus O Watson
Journal:  J Hypertens       Date:  2017-03       Impact factor: 4.844

3.  Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension.

Authors:  Jalil Belghazi; Ramzi N El Feghali; Thérèse Moussalem; Maya Rejdych; Roland G Asmar
Journal:  Vasc Health Risk Manag       Date:  2007
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.