The heart-pump interaction: effects of a microaxial blood pump.

J Stoliński1, C Rosenbaum, W Flameng, B Meyns.   

Abstract

BACKGROUND: When we use rotary blood pumps as an assist device, an interaction takes place between the pump performance and the native heart function (native heart influences pump performance and vice versa). The interaction between native heart and rotary blood pump can be useful to predict recovery of the failing heart.
METHODS: The rotary blood pumps used were microaxial catheter-mounted pumps with an external diameter of 6.4 mm (Impella, Aachen, Germany). The pump-heart interaction was studied in five juvenile sheep with a mean body weight of 68.5 +/- 8.7 kg. The pumps were introduced via the left carotid artery and placed in transvalvular aortic position. Recorded parameters were pump speed (rpm), generated flow (L/min) and differential pressure (mm Hg) obtained at high frequency rate of data recordings (25 sets of data per second). This allowed continuous analysis of the pump performance during cardiac cycle. Under clinical conditions the interaction was studied in a 60-year-old male, in whom the device was applied due to postcardiotomy heart failure after myocardial infarction.
RESULTS: Heart-pump interaction was analyzed based on pump flow differential pressure. This relationship, analyzed continuously during cardiac cycle, presents as a loop. The dynamic contribution of the heart to the flow generated by the pump leads to continuous fluctuation in the pressure head and the creation of hysteresis. The improved function of the failing heart under clinical conditions after seven days of mechanical support was expressed by: increased hysteresis of the loop caused by increased gradient of flow generated during cardiac cycle, a more pronounced venticular ejection phase that indicates more dynamic heart contribution to the generated flow, and finally increased gradient of the differential pressure during cardiac cycle, caused predominantly by increased aortic pressure and decreased left ventricle pressure during diastolic phase.
CONCLUSIONS: The heart-pump interaction based on the pump flow-differential pressure relationship can be useful in predicting the possibility to wean the patient from the device.

Entities:  

Mesh:

Year:  2002        PMID: 12487396     DOI: 10.1177/039139880202501107

Source DB:  PubMed          Journal:  Int J Artif Organs        ISSN: 0391-3988            Impact factor:   1.595


  2 in total

1.  Dynamic Modulation of Device-Arterial Coupling to Determine Cardiac Output and Vascular Resistance.

Authors:  Steven P Keller; Brian Y Chang; Qing Tan; Zhengyang Zhang; Ahmad El Katerji; Elazer R Edelman
Journal:  Ann Biomed Eng       Date:  2020-04-13       Impact factor: 3.934

2.  Hysteretic device characteristics indicate cardiac contractile state for guiding mechanical circulatory support device use.

Authors:  Brian Y Chang; Zhengyang Zhang; Steven P Keller; Elazer R Edelman; Kimberly Feng; Noam Josephy
Journal:  Intensive Care Med Exp       Date:  2021-12-20
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.