Literature DB >> 12483088

Learning effect and measurement variability in frequency-doubling technology perimetry in chronic open-angle glaucoma.

Hiroshi Matsuo1, Goji Tomita, Yasuyuki Suzuki, Makoto Araie.   

Abstract

PURPOSE: To prospectively study the learning effects and measurement variability in frequency-doubling technology perimetry in glaucoma patients who were well practiced in testing with an automated static perimeter. PATIENTS AND METHODS: Patients with clinically stable chronic open-angle glaucoma (n = 44) who had undergone repeated perimetric testing using a 30-2 full-threshold program with a Humphrey perimeter with reliable results were enrolled in the study. The authors tested each patient using frequency-doubling technology perimetry. After initial examination with the C-20-1 screening program, the N-30 full-threshold program was performed three times at 1- to 3-month intervals.
RESULTS: Perimetric results of the N-30 program improved with repeated testing in both eyes between the first and second visits. The average of mean deviation (MD) of the N-30 program significantly improved by 0.6 dB ( < 0.001) in the right eye and 0.8 dB ( < 0.002) in the left eye between the first and second sessions. The sensitivity in 4 of 19 test points (right eye) and 5 of 19 points (left eye) improved significantly at the second sessions, but not the third session. There were no significant differences in the FDT results between the second and third sessions in both eyes. The standard deviation of the mean MD value for the three sessions ranged from 3.8 to 4.0 dB in both eyes.
CONCLUSION: Baseline measurements for perimetric follow-up with the N-30 full-threshold program in patients with glaucoma should be repeated at least twice after testing with the C-20-1 screening program.

Entities:  

Mesh:

Year:  2002        PMID: 12483088     DOI: 10.1097/00061198-200212000-00002

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  5 in total

1.  Recovery of reliability by retest after a 5-minute interval in frequency doubling technology perimetry.

Authors:  Hidetoshi Tsukamoto; Satoshi Mukai; Aiko Iwase; Hiromu K Mishima
Journal:  Jpn J Ophthalmol       Date:  2006 Jul-Aug       Impact factor: 2.447

Review 2.  Effect of cataract extraction on frequency doubling technology perimetry in patients with glaucoma.

Authors:  M A R Siddiqui; A Azuara-Blanco; S Neville
Journal:  Br J Ophthalmol       Date:  2005-12       Impact factor: 4.638

3.  Glaucoma detection with matrix and standard achromatic perimetry.

Authors:  Zvia Burgansky-Eliash; Gadi Wollstein; Avni Patel; Richard A Bilonick; Hiroshi Ishikawa; Larry Kagemann; William D Dilworth; Joel S Schuman
Journal:  Br J Ophthalmol       Date:  2007-01-10       Impact factor: 4.638

4.  Large-scale survey of unselected automated visual fields in a major reading center: patterns and data analysis.

Authors:  Lilly Zborowski-Naveh; Rita Ehrlich; Moshe Luski; Dov Weinberger; Mona Boaz; Dan D Gaton
Journal:  Scientifica (Cairo)       Date:  2012-05-20

5.  Changes in Visual Field Indices after Pterygium Surgery.

Authors:  Fikret Ucar; Halil Ibrahim Yener; Servet Cetinkaya; Huseyina Ture
Journal:  Beyoglu Eye J       Date:  2021-12-17
  5 in total