Conversion factor for comparison of data from Humphrey and Medmont automated perimeters.

David Pye1, Peter Herse, Ha Nguyen, Lan Vuong, Quoc Pham.   

Abstract

Background: Clinical experience has shown that the sensitivity indices reported by the Humphrey Field Analyser (HFA) are generally higher than those given by the Medmont Automated Perimeter (M600). It is the purpose of this paper to determine a conversion factor for the two perimeters and to confirm this prediction using clinical data. Theory predicted that HFA(sensitivity) - 5 dB = M600(sensitivity).
Methods: Sensitivity versus eccentricity profiles were measured over the central visual field on 10 young subjects using both perimeters.
Results: Both the HFA and the M600 operate within the realms of the Weber law and measure similar Weber fractions. The sensitivity profiles had similar slopes (about -0.2 dB/degree) and were separated by about six decibels with the HFA reporting higher sensitivity values. This result confirmed the theoretical prediction.
Conclusion: The difference in threshold sensitivities between the two perimeters is a result of differences in scaling factors and instrument luminances. A suggested clinical conversion factor is to subtract 5 dB from the HFA data to approximate those of the M600.

Entities:  

Year:  1999        PMID: 12482302     DOI: 10.1111/j.1444-0938.1999.tb06779.x

Source DB:  PubMed          Journal:  Clin Exp Optom        ISSN: 0816-4622            Impact factor:   2.742


  3 in total

1.  A comparison of perimetric results with the Medmont and Humphrey perimeters.

Authors:  J Landers; A Sharma; I Goldberg; S Graham
Journal:  Br J Ophthalmol       Date:  2003-06       Impact factor: 4.638

2.  A comparison of global indices between the Medmont Automated Perimeter and the Humphrey Field Analyzer.

Authors:  John Landers; Alok Sharma; Ivan Goldberg; Stuart Graham
Journal:  Br J Ophthalmol       Date:  2007-03-27       Impact factor: 4.638

Review 3.  The value of visual field testing in the era of advanced imaging: clinical and psychophysical perspectives.

Authors:  Jack Phu; Sieu K Khuu; Michael Yapp; Nagi Assaad; Michael P Hennessy; Michael Kalloniatis
Journal:  Clin Exp Optom       Date:  2017-06-22       Impact factor: 2.742

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.