Measurements of X ray absorbed doses to dental patients in two dental X ray units in Nigeria.

F O Ogundaret1, O M Oni, F A Balogun.   

Abstract

Measurements of absorbed doses from radiographic examinations to various anatomical sites in the head and neck of patients with an average age of 45 years using intra-oral dental radiography have been carried out. LiF (TLD-100) dosemeters were used for the measurements of the absorbed dose. The measured absorbed doses to the various anatomical sites in the two units are reported, discussed and compared with results from the literature. Quality control measurements were also performed using a Victoreen quality control test device on the X ray units. The tube voltage accuracies for the two units were found to be within acceptable limits (less than +/- 10%). On the other hand the exposure time accuracies for these units have large deviations (>20%). These results and those that have been reported in the literature may be an indication that high patient doses are common in most dental X ray centres and countries. As a result of this, regular compliance and performance checks of dental diagnostic X ray equipment are essential in order to ensure proper performance and to minimise unnecessary patient and operator doses.

Entities:  

Mesh:

Year:  2002        PMID: 12474946     DOI: 10.1093/oxfordjournals.rpd.a006105

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  1 in total

1.  Influence of lead apron shielding on absorbed doses from panoramic radiography.

Authors:  D Rottke; L Grossekettler; K Sawada; P Poxleitner; D Schulze
Journal:  Dentomaxillofac Radiol       Date:  2013-10-30       Impact factor: 2.419

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.