An ex-vivo evaluation of vascular catheters for continuous hemofiltration.

Han Khim Tan1, Nicholas Bridge, Ian Baldwin, Rinaldo Bellomo.   

Abstract

OBJECTIVES: To measure outflow and inflow hydraulic resistance in double-lumen catheters used for hemofiltration under standardized laboratory conditions.
SETTING: ICU Laboratory of tertiary unit.
METHODS: Heparinized spent red cells diluted in polygeline solution to a constant hematocrit of 32% at 37 degrees C were pumped using a standard Prisma M60 circuit through several hemofiltration catheters. Blood pump speed was increased and decreased in steps of 30mL/min (30, 60, 90, 120, 150, and 180 mL/min) and catheter outflow and inflow pressures recorded and used to define the pressure flow relationship (line of hydraulic resistance) for each.
RESULTS: Double-lumen catheters posed different resistances to outflow or inflow. Among the < 15 cm long catheters, the 11.5 Fr Quinton-Mahurkar (0.56 mmHg/mL/min) catheter offered the least resistance to outflow, while the Medcomp 11.5 Fr catheter offered the least resistance to inflow (0.78 mmHg/mL/min). Among the >19 cm long catheters, the 13.5 Fr Vascath Niagara catheter showed the lowest blood flow resistance to both outflow (0.63 mmHg/ mL/min) and inflow (0.83 mmHg/mL/min). Longer catheters did not pose statistically greater resistance to both outflow and inflow. Resistance to inflow was consistently greater than resistance to outflow (p = 0.003). Overall, the Prisma M60 blood circuit alone accounted for 40% of the total extracorporeal circuit blood flow resistance.
CONCLUSIONS: Proprietary hemofiltration catheters have variable resistance to blood flow under standard ex-vivo conditions. This ex-vivo information might be useful to clinicians in guiding their choice of catheters for clinical use.

Entities:  

Mesh:

Year:  2002        PMID: 12472198     DOI: 10.1081/jdi-120015678

Source DB:  PubMed          Journal:  Ren Fail        ISSN: 0886-022X            Impact factor:   2.606


  1 in total

1.  Dual-lumen catheters for continuous venovenous hemofiltration: limits for blood delivery via femoral vein access and a potential alternative in an experimental setting in anesthetized pigs.

Authors:  Juliane K Unger; Klaus Pietzner; Roland C Francis; Juergen Birnbaum; Marc Michael Theisen; Arne-Joern Lemke; Stefan M Niehues
Journal:  Crit Care       Date:  2007       Impact factor: 9.097

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.