Literature DB >> 12469043

Multicenter trial of the PowerLink bifurcated system for endovascular aortic aneurysm repair.

Jeffrey P Carpenter1.   

Abstract

PURPOSE: The purpose of this study was to assess the results of abdominal aortic aneurysm repair with the PowerLink bifurcated endovascular graft (Endologix, Inc, Irvine, Calif).
METHOD: Twelve centers used the PowerLink bifurcated system for elective endovascular aneurysm repair in 118 patients recruited during a 16-month interval and followed for a 25-month interval (mean follow-up, 16 months) as part of a pivotal US Food and Drug Administration trial. Stent grafts were oversized by 10% to 20% relative to computed tomographic scan-based diameter measurements. All repairs were performed in the operating room through one surgically exposed femoral artery and a contralateral 9F sheath percutaneously placed. Results were assessed with contrast-enhanced computed tomography and plain abdominal radiography at 1, 6, and 12 months after surgery.
RESULTS: Three failed insertions and one late conversion for endoleak remediation occurred, resulting in four conversions (3.3%) to open surgery. Of the failed insertions, two were from a faulty delivery system design, which was corrected. No failures occurred after the modification. One perioperative death (0.8%) occurred that was not device related. Eight late deaths were from unrelated causes, and one was from complications after reoperation for treatment of an endoleak. Endoleaks were noted in 19 patients (16%) at the time of the endograft procedure: 12 resolved spontaneously, four resolved with secondary interventions (three type I, one type II), and three underwent observation, yielding a 30-day endoleak rate of 5.9%. Two graft limb thromboses (0.8%) were seen. One graft migration (0.8%) was of no clinical significance. No ruptures or wire fractures were found. The mean aneurysm diameter was reduced from 51 mm (preoperative) to 45 mm (12 months; P <.0001).
CONCLUSION: The PowerLink system appears to be safe and effectively protects patients from abdominal aortic aneurysm rupture over the short to medium term. The low endoleak rate is superior to that reported for other devices. The graft and stent materials have thus far been free from failure and fatigue. The sutureless stent and endoskeleton design confer a number of unique advantages and challenges. Careful follow-up over the longer term is necessary to assure the durability of these results.

Entities:  

Mesh:

Year:  2002        PMID: 12469043     DOI: 10.1067/mva.2002.129641

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  4 in total

1.  Utility of aortic cuffs in converting initially ineligible patients due to unfavorable neck anatomy into successful candidates for endovascular aortic aneurysm repair: A Case Series.

Authors:  Omer Awan; Mark Garcia; Yousaf Awan; Mandip Gakhal; Moonjohn Kim; Bogdan Iliescu; Eliot Siegel
Journal:  J Radiol Case Rep       Date:  2010-03-01

2.  Lifeline registry of endovascular aneurysm repair: open repair surgical controls in clinical trials.

Authors:  Robert M Zwolak; Anton N Sidawy; Roy K Greenberg; Marc L Schermerhorn; Rebecca J Shackelton; Flora S Siami
Journal:  J Vasc Surg       Date:  2008-07-23       Impact factor: 4.268

3.  Outcomes of Unibody Bifurcated Endograft and Aortobifemoral Bypass for Aortoiliac Occlusive Disease.

Authors:  Baker Ghoneim; Mohamed Elsherif; Mohamed Elsharkawi; Yogesh Acharya; Niamh Hynes; Wael Tawfick; Sherif Sultan
Journal:  Vasc Specialist Int       Date:  2020-12-31

4.  Midterm outcomes of 455 patients receiving the AFX2 endovascular graft for the treatment of abdominal aortic aneurysm: A retrospective multi-center analysis.

Authors:  Raymond Vetsch; Harvey E Garrett; Christopher L Stout; Alan R Wladis; Matt Thompson; Joseph V Lombardi
Journal:  PLoS One       Date:  2021-12-31       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.