Literature DB >> 12461373

Late dislocation of rotating platform in New Jersey Low-Contact Stress knee prosthesis.

Chun-Hsiung Huang1, Hon-Ming Ma, Jiann-Jong Liau, Fang-Yuan Ho, Cheng-Kung Cheng.   

Abstract

Five patients with late rotational dislocation of the rotating platform bearing in the New Jersey Low-Contact Stress total knee arthroplasty are reported. The prostheses had functioned well for 8 to 12 years before failure. Preoperative radiographs showed asymmetric femorotibial joint spaces. Entrapment of the dislocated bearing in three patients and spontaneous reduction of the dislocated bearing in another two patients were seen at revision. Femorotibial ligamentous instability was found after reduction. The retrieved polyethylene bearings showed advanced wear and cold flow deformities and the thickness was reduced. The revision arthroplasty was accomplished by replacement with a thicker bearing element. Progressive femorotibial ligament laxity and reduction of the thickness of polyethylene with wearing break down the originally well-balanced soft tissue tension of the knee. The rotational degree of the rotating platform bearing is unrestricted, which may result in late dislocation. Polyethylene wear is unavoidable in knee prostheses using metal contact with polyethylene even with a mobile-bearing design. Efforts to reduce polyethylene wear are mandatory.

Entities:  

Mesh:

Year:  2002        PMID: 12461373     DOI: 10.1097/00003086-200212000-00024

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  7 in total

Review 1.  [From tibiofemoral instability to dislocation in total knee arthroplasty].

Authors:  M Pietsch; S Hofmann
Journal:  Orthopade       Date:  2007-10       Impact factor: 1.087

2.  Fixed or mobile-bearing total knee arthroplasty.

Authors:  Chun-Hsiung Huang; Jiann-Jong Liau; Cheng-Kung Cheng
Journal:  J Orthop Surg Res       Date:  2007-01-05       Impact factor: 2.359

3.  What have we learned from 100% success of press fit condylar rotating platform posterior stabilized knees?: A 5-10 years followup by a nondesigner.

Authors:  Shrinand V Vaidya; Siddharth Virani; Rajendra Phunde; Abhishek Mahajan
Journal:  Indian J Orthop       Date:  2016 Nov-Dec       Impact factor: 1.251

4.  Migration and clinical outcome of mobile-bearing versus fixed-bearing single-radius total knee arthroplasty.

Authors:  Koen T Van Hamersveld; Perla J Marang-Van De Mheen; Huub J L Van Der Heide; Henrica M J Van Der Linden-Van Der Zwaag; Edward R Valstar; Rob G H H Nelissen
Journal:  Acta Orthop       Date:  2018-02-16       Impact factor: 3.717

5.  Mid- term results of stryker® scorpio plus mobile bearing total knee arthroplasty.

Authors:  Hideo Kobayashi; Naoto Mitsugi; Yuichi Mochida; Naoya Taki; Yasushi Akamatsu; Masato Aratake; Hirohiko Ota; Katsushi Ishii; Kengo Harigane; Taichi Ideno; Tomoyuki Saito
Journal:  Sports Med Arthrosc Rehabil Ther Technol       Date:  2012-10-18

Review 6.  Biomechanical considerations in the design of high-flexion total knee replacements.

Authors:  Cheng-Kung Cheng; Colin J McClean; Yu-Shu Lai; Wen-Chuan Chen; Chang-Hung Huang; Kun-Jhih Lin; Chia-Ming Chang
Journal:  ScientificWorldJournal       Date:  2014-05-06

7.  MOBILE TIBIAL POLYETHYLENE BEARING IN TOTAL KNEE ARTHROPLASTY.

Authors:  Hugo Alexandre de Araújo Barros Cobra; Idemar Monteiro da Palma
Journal:  Rev Bras Ortop       Date:  2015-12-07
  7 in total