Optimization of blood oxygenation level-dependent sensitivity in magnetic resonance imaging using intermolecular double-quantum coherence.

Jianhui Zhong1, Zhong Chen, W Edmund Kwok, Scott Kennedy, Zhiyang You.   

Abstract

PURPOSE: To optimize timing parameters in an intermolecular double-quantum coherence (iDQC) imaging pulse sequence for overall image signal-to-noise ratio (SNR) and blood oxygenation level-dependent (BOLD) sensitivity for brain functional imaging.
MATERIAL AND METHODS: Fresh human blood was measured under different oxygenation conditions, and human brain functional magnetic resonance (fMR) images in three normal volunteers were obtained, using iDQC techniques at 1.5 T. The dependence of SNR and BOLD sensitivity was measured as a function of time delays after the iDQC evolution period.
RESULTS: A time delay after the iDQC evolution period tau can be adjusted either to refocus the dephasing accumulated during tau, thus increasing SNR, with full rephasing occurring at delay = +/-2tau (for iDQC order n = +/-2), or to enhance BOLD effects with consequent reduced image SNR at delay = 0.
CONCLUSION: Image SNR and BOLD sensitivity often impose different requirements for iDQC image sequence design and timing parameter selections. It is therefore important to select properly relevant parameters for different applications. Copyright 2002 Wiley-Liss, Inc.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12451587     DOI: 10.1002/jmri.10208

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  2 in total

1.  Functional MRI at 3T using intermolecular double-quantum coherence (iDQC) with spin-echo (SE) acquisitions.

Authors:  T Gu; S D Kennedy; Z Chen; K A Schneider; J Zhong
Journal:  MAGMA       Date:  2007-12-05       Impact factor: 2.310

2.  Numerical simulations of motion-insensitive diffusion imaging based on the distant dipolar field effects.

Authors:  Tao Lin; Huijun Sun; Zhong Chen; Rongyi You; Jianhui Zhong
Journal:  Magn Reson Imaging       Date:  2007-06-13       Impact factor: 2.546

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.