Literature DB >> 12450718

Reliability and validity issues of laser fluorescence measurements in occlusal caries diagnosis.

Hanna M Alwas-Danowska1, Alphons J M Plasschaert, Stanislaw Suliborski, Emiel H Verdonschot.   

Abstract

OBJECTIVES: Recently, a device for detecting occlusal caries lesions based on laser fluorescence measurements (DIAGNOdent) has been introduced. The reliability (reproducibility) and validity of this laser-fluorescence device were investigated.
METHODS: In the in vitro part of the study, 49 permanent molars were measured by two observers using two DIAGNOdent devices. In addition, visual inspection was performed. The teeth were sectioned to measure the actual depth and area of the caries lesions. In the in vivo part of the study, 45 sites at the occlusal surfaces of permanent molar teeth in 13 patients were measured by two observers using two DIAGNOdent devices, manufactured in two successive years.
RESULTS: The reliability of both DIAGNOdent devices expressed in the intraclass correlation coefficient (ICC=0.90 for observer 1, and 0.88 for observer 2) was high, and so was the interobserver reliability (ICC=0.85 for the 1998-device, ICC=0.90 for the 1999-device). The correlation between DIAGNOdent measurements and the actual depth of the caries lesions was lower than that of visual inspection by one observer, yet higher than that of a second observer. The specificities of visual inspection (0.94 and 0.88) were higher than those of the DIAGNOdent devices, whereas the sensitivities of the DIAGNOdent devices (0.93-1.00) exceeded those of visual inspection The validity of the DIAGNOdent, expressed as the area under the receiver operating characteristic curve, was not statistically significantly different from that of visual inspection (P>0.05). The DIAGNOdent measurements showed a higher association with the enamel part of the caries lesion than with the dentinal part.
CONCLUSIONS: The DIAGNOdent is suitable for monitoring small caries lesions. Because of the high reproducibility of DIAGNOdent devices produced in successive years, a DIAGNOdent device may be replaced by a new one. Due to the higher specificity, visual inspection should be preferred to diagnose occlusal caries lesions in populations with a low caries prevalence.

Entities:  

Mesh:

Year:  2002        PMID: 12450718     DOI: 10.1016/s0300-5712(02)00015-5

Source DB:  PubMed          Journal:  J Dent        ISSN: 0300-5712            Impact factor:   4.379


  20 in total

1.  Laser fluorescence of dentin caries covered with a novel nano-filled sealant.

Authors:  Andreas Braun; Christian Beisel; Olivier Brede; Felix Krause
Journal:  Lasers Med Sci       Date:  2012-03-30       Impact factor: 3.161

2.  In vivo comparison of laser fluorescence measurements with conventional methods for occlusal caries detection.

Authors:  Meryem Toraman Alkurt; Ilkay Peker; Hacer Deniz Arisu; Oya Bala; Bülent Altunkaynak
Journal:  Lasers Med Sci       Date:  2007-09-21       Impact factor: 3.161

3.  The effect of pit and fissure sealants on the detection of occlusal caries in vitro.

Authors:  D J Manton; L B Messer
Journal:  Eur Arch Paediatr Dent       Date:  2007-03

4.  Histological validation of a laser fluorescence device for occlusal caries detection in primary molars.

Authors:  D Apostolopoulou; P Lagouvardos; K Kavvadia; L Papagiannoulis
Journal:  Eur Arch Paediatr Dent       Date:  2009-11

5.  Combined validity of DIAGNOdent™ and visual examination for in vitro detection of occlusal caries in primary molars.

Authors:  Katerina Kavvadia; Panagiotis Lagouvardos; Daphne Apostolopoulou
Journal:  Lasers Med Sci       Date:  2011-01-18       Impact factor: 3.161

6.  Performance differences in the detection of subgingival calculus by laser fluorescence devices.

Authors:  Fardad Shakibaie; Laurence J Walsh
Journal:  Lasers Med Sci       Date:  2015-09-22       Impact factor: 3.161

7.  The efficiency of operating microscope compared with unaided visual examination, conventional and digital intraoral radiography for proximal caries detection.

Authors:  Ilkay Peker; Meryem Toraman Alkurt; Oya Bala; Bulent Altunkaynak
Journal:  Int J Dent       Date:  2009-01-29

8.  In situ and in vitro comparison of laser fluorescence with visual inspection in detecting occlusal caries lesions.

Authors:  Andréia Bolzan de Paula; Juliana Alvares Duarte Bonini Campos; Michele Baffi Diniz; Josimeri Hebling; Jonas Almeida Rodrigues
Journal:  Lasers Med Sci       Date:  2009-09-26       Impact factor: 3.161

9.  Caries diagnosis using light fluorescence devices in comparison with traditional visual and tactile evaluation: a prospective study in 152 patients.

Authors:  María Melo; Agustín Pascual; Isabel Camps; Ángel Del Campo; Javier Ata-Ali
Journal:  Odontology       Date:  2016-09-21       Impact factor: 2.634

10.  Comparison between a laser fluorescence device and visual examination in the detection of occlusal caries in children.

Authors:  Chaza Kouchaji
Journal:  Saudi Dent J       Date:  2012-08-17
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.