Literature DB >> 12433038

A rapid improved method for gamma-spectrometric determination of 202Tl impurities in [201Tl]-labelled radiopharmaceuticals.

Mauro Bonardi1, Flavia Groppi, Claudio Birattari.   

Abstract

Despite the cyclotron production method and the efficiency of the radiochemical procedures adopted, the long-lived radio-isotopic impurity 202Tl is always present in [201Tl]-labelled radio-pharmaceuticals, together with other short-lived impurities like, 200Tl. Rapid determination of the 202Tl impurity, can be achieved using HPGe gamma spectrometry and a detector shielded by a 5 mm thick envelope of lead. In this way, dead-time correction errors, Compton and X-ray background, are very efficiently avoided and suppressed. The same method could be applied routinely in nuclear medicine, to determine the radioisotopic purity of 201Tl by means of an ionisation chamber dose calibrator.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12433038     DOI: 10.1016/s0969-8043(02)00179-3

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  1 in total

1.  Degradation of myocardial perfusion SPECT images caused by contaminants in thallous (201Tl) chloride.

Authors:  Steven G Staelens; Tim C de Wit; Ignace A Lemahieu; Freek J Beekman
Journal:  Eur J Nucl Med Mol Imaging       Date:  2008-01-25       Impact factor: 9.236