Comparative evaluation of JPEG and JPEG2000 compression in quantitative digital subtraction radiography.

A Fidler1, B Likar, F Pernus, U Skaleric.   

Abstract

OBJECTIVES: The aim of this in vitro study was to compare the impact of JPEG and the novel JPEG2000 compression standard on quantitative digital subtraction radiography (DSR) and to determine the acceptable JPEG2000 compression ratios for DSR.
METHODS: Nine dry pig mandible sections were radiographed three times ('Baseline', 'No change', and 'Gain') with standardized projection geometry. Bone gain was simulated by adding artificial bone chips (1, 4 and 15 mg). Images were registered, compressed by JPEG and JPEG2000 using compression ratios (CR) of 1 : 7, 1 : 16, 1 : 22, and 1 : 31, and then subtracted. Image distortion was assessed objectively by calculating average pixel error and peak signal to noise ratio. No change areas in compressed and subtracted 'No change-Baseline' images and bone gain volumes in compressed and subtracted 'Gain-Baseline' images were calculated for both compression standards and compared.
RESULTS: JPEG introduced less distortion at low CRs, while JPEG2000 was superior at higher CRs. At CR of 1 : 7, no significant difference between JPEG and JPEG2000 was found. JPEG2000 yielded better results for no change measurements at higher CRs. Volumes of simulated bone gain were overestimated when JPEG and underestimated when JPEG2000 compression was used.
CONCLUSIONS: At CR of 1 : 7 JPEG and JPEG2000 performed similarly, which indicates that CR of 1:7 in JPEG2000 can be used for DSR if images are registered before compression. At higher CRs, JPEG2000 is superior to JPEG but image distortions are too high for reliable quantitative DSR.

Entities:  

Mesh:

Year:  2002        PMID: 12424637     DOI: 10.1038/sj.dmfr.4600724

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  3 in total

1.  Fractal analysis of periapical bone from lossy compressed radiographs: a comparison of two lossy compression methods.

Authors:  B Güniz Baksi; Aleš Fidler
Journal:  J Digit Imaging       Date:  2011-12       Impact factor: 4.056

2.  Virtual slide telepathology workstation of the future: lessons learned from teleradiology.

Authors:  Elizabeth A Krupinski
Journal:  Hum Pathol       Date:  2009-06-24       Impact factor: 3.466

3.  Effect of image compression of direct digital lateral cephalograms on the identification of cephalometric points.

Authors:  Sima Saghaie; Roshanak Ghaffari
Journal:  Dent Res J (Isfahan)       Date:  2014-01
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.