Acetabular component migration in total hip arthroplasty using CT and a semiautomated program for volume merging.

L Olivecrona1, J Crafoord, H Olivecrona, M E Noz, G Q Maguire, M P Zeleznik, L Svensson, L Weidenhielm.   

Abstract

PURPOSE: To develop a non-invasive method for detection of acetabular cup migration after total hip arthroplasty (THA) with a higher degree of accuracy than routine plain radiography.
MATERIAL AND METHODS: Two CT examinations, 10 min apart, were obtained from each of 10 patients that had undergone THA. Using an in-house developed semiautomated program for volume merging, the pelves in the two examinations were fused and the acetabular cup was visually and numerically evaluated to test the method's accuracy in detecting migration.
RESULTS: In the visual evaluation of the best match a 1-mm translation of the cup was detectable. The numerical evaluation, comparing landmarks placed in the images of the acetabular cup and the head of the femur component in the two examinations, showed the mean difference in orientation of acetabular axes to be 2.5 degrees, the mean distance between centre of cup face to be 2.5 mm and the mean distance between centre of the head of the prosthetic femoral component to be 1 mm.
CONCLUSION: This method has a significantly higher accuracy than routine plain radiography in detecting acetabular cup migration and could be used in clinical practice. It gives both a visual and a numerical correlate to migration.

Entities:  

Mesh:

Year:  2002        PMID: 12423464

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  14 in total

1.  Three-dimensional movements of the lumbar spine facet joints and segmental movements: in vivo examinations of normal subjects with a new non-invasive method.

Authors:  P Svedmark; T Tullberg; M E Noz; G Q Maguire; M P Zeleznik; L Weidenhielm; G Nemeth; H Olivecrona
Journal:  Eur Spine J       Date:  2011-09-01       Impact factor: 3.134

2.  Validation of a 3D CT method for measurement of linear wear of acetabular cups.

Authors:  Anneli Jedenmalm; Fritjof Nilsson; Marilyn E Noz; Douglas D Green; Ulf W Gedde; Ian C Clarke; Andreas Stark; Gerald Q Maguire; Michael P Zeleznik; Henrik Olivecrona
Journal:  Acta Orthop       Date:  2011-02-01       Impact factor: 3.717

3.  Using PET/CT Bone Scan Dynamic Data to Evaluate Tibia Remodeling When a Taylor Spatial Frame Is Used: Short and Longer Term Differences.

Authors:  Henrik Lundblad; Gerald Q Maguire; Charlotte Karlsson-Thur; Cathrine Jonsson; Marilyn E Noz; Michael P Zeleznik; Hans Jacobsson; Lars Weidenhielm
Journal:  Biomed Res Int       Date:  2015-09-07       Impact factor: 3.411

4.  In vivo and ex vivo measurement of polyethylene wear in total hip arthroplasty: comparison of measurements using a CT algorithm, a coordinate-measuring machine, and a micrometer.

Authors:  Dov Goldvasser; Viktor J Hansen; Marilyn E Noz; Gerald Q Maguire; Michael P Zeleznik; Henrik Olivecrona; Charles R Bragdon; Lars Weidenhielm; Henrik Malchau
Journal:  Acta Orthop       Date:  2014-04-23       Impact factor: 3.717

5.  Accuracy and Precision of Three-Dimensional Low Dose CT Compared to Standard RSA in Acetabular Cups: An Experimental Study.

Authors:  Cyrus Brodén; Henrik Olivecrona; Gerald Q Maguire; Marilyn E Noz; Michael P Zeleznik; Olof Sköldenberg
Journal:  Biomed Res Int       Date:  2016-07-10       Impact factor: 3.411

6.  A CT method for following patients with both prosthetic replacement and implanted tantalum beads: preliminary analysis with a pelvic model and in seven patients.

Authors:  Henrik Olivecrona; Gerald Q Maguire; Marilyn E Noz; Michael P Zeleznik; Uldis Kesteris; Lars Weidenhielm
Journal:  J Orthop Surg Res       Date:  2016-02-24       Impact factor: 2.359

7.  Three-dimensional analysis of cervical spine segmental motion in rotation.

Authors:  Xiong Zhao; Zi-Xiang Wu; Bao-Jun Han; Ya-Bo Yan; Yang Zhang; Wei Lei
Journal:  Arch Med Sci       Date:  2013-05-27       Impact factor: 3.318

8.  A new automated way to measure polyethylene wear in THA using a high resolution CT scanner: method and analysis.

Authors:  Gerald Q Maguire; Marilyn E Noz; Henrik Olivecrona; Michael P Zeleznik; Lars Weidenhielm
Journal:  ScientificWorldJournal       Date:  2014-01-22

9.  Can Na18F PET/CT be used to study bone remodeling in the tibia when patients are being treated with a Taylor Spatial Frame?

Authors:  Henrik Lundblad; Gerald Q Maguire; Henrik Olivecrona; Cathrine Jonsson; Hans Jacobsson; Marilyn E Noz; Michael P Zeleznik; Lars Weidenhielm; Anders Sundin
Journal:  ScientificWorldJournal       Date:  2014-03-19

10.  A New CT Method for Assessing 3D Movements in Lumbar Facet Joints and Vertebrae in Patients before and after TDR.

Authors:  Per Svedmark; Svante Berg; Marilyn E Noz; Gerald Q Maguire; Michael P Zeleznik; Lars Weidenhielm; Gunnar Nemeth; Henrik Olivecrona
Journal:  Biomed Res Int       Date:  2015-10-26       Impact factor: 3.411

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.