| Literature DB >> 12418750 |
Dirk Vollmann1, Lars Lüthje, Georg Görtler, Christina Unterberg.
Abstract
Oversensing of intracardiac signals or myopotentials may cause inappropriate ICD therapy. Reports on far-field sensing of atrial signals are rare, and inappropriate ICD therapy due to oversensing of atrial fibrillation has not yet been described. This report presents a patient with a triple chamber ICD and a history of His-bundle ablation who experienced asystolic ventricular pauses and inappropriate detection of ventricular fibrillation due to far-field oversensing of atrial fibrillation. Several factors contributed to the complication, which resolved after reduction of the ventricular sensitivity.Entities:
Mesh:
Year: 2002 PMID: 12418750 DOI: 10.1046/j.1460-9592.2002.01513.x
Source DB: PubMed Journal: Pacing Clin Electrophysiol ISSN: 0147-8389 Impact factor: 1.976