Multiple gradient echo sequence optimized for rapid, single-scan mapping of R(2)(*) at high B0.

Jim M Wild1, W R Wayne Martin, Peter S Allen.   

Abstract

A multiple-gradient-echo sequence is proposed for accurately mapping R(2)(*) in the presence of in-slice macroscopic susceptibility gradients. In-slice signal loss caused by background macroscopic susceptibility gradients is mitigated by combining three successive gradient-echo images whose slice refocus gradients are successively incremented. The optimum incrementation of slice-refocusing gradients was determined by numerical simulation. By repeating further cycles of three images in the same sequence, artifact-compensated data spanning a range of echo times (TEs) was acquired leading to single-scan, R(2) (*) maps that are quantitatively reflective of microscopic field inhomogeneities. The performance of the sequence was demonstrated at 3.0T, first with a doped aqueous phantom, and then on the head of a normal volunteer. That performance is compared quantitatively with previously published work. Copyright 2002 Wiley-Liss, Inc.

Mesh:

Year:  2002        PMID: 12418002     DOI: 10.1002/mrm.10291

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  6 in total

1.  R*(2) mapping in the presence of macroscopic B₀ field variations.

Authors:  Diego Hernando; Karl K Vigen; Ann Shimakawa; Scott B Reeder
Journal:  Magn Reson Med       Date:  2011-12-09       Impact factor: 4.668

2.  An efficient automated z-shim based method to correct through-slice signal loss in EPI at 3T.

Authors:  Helen Marshall; Joseph V Hajnal; Jane E Warren; Richard J Wise; David J Larkman
Journal:  MAGMA       Date:  2009-02-24       Impact factor: 2.310

3.  Susceptibility gradient quantization by MRI signal response mapping (SIRMA) to dephaser.

Authors:  F Franconi; C Chapon; J J Le Jeune; P Richomme; L Lemaire
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

4.  Voxel spread function method for correction of magnetic field inhomogeneity effects in quantitative gradient-echo-based MRI.

Authors:  Dmitriy A Yablonskiy; Alexander L Sukstanskii; Jie Luo; Xiaoqi Wang
Journal:  Magn Reson Med       Date:  2012-12-11       Impact factor: 4.668

5.  Adaptive slice-specific z-shimming for 2D spoiled gradient-echo sequences.

Authors:  Martin Soellradl; Johannes Strasser; Andreas Lesch; Rudolf Stollberger; Stefan Ropele; Christian Langkammer
Journal:  Magn Reson Med       Date:  2020-09-10       Impact factor: 4.668

6.  Assessment and correction of macroscopic field variations in 2D spoiled gradient-echo sequences.

Authors:  Martin Soellradl; Andreas Lesch; Johannes Strasser; Lukas Pirpamer; Rudolf Stollberger; Stefan Ropele; Christian Langkammer
Journal:  Magn Reson Med       Date:  2019-12-23       Impact factor: 4.668

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.