| Literature DB >> 12416880 |
Sarah Pelletier1, Charles A Lucy.
Abstract
A fluorescent post-column reaction detection scheme has been devised for selective determination of thiols. The post-column reagent is 40 microM Cd2+ and 100 microM 8-hydroxyquinoline-5-sulfonic acid (HQS) in non-complexing buffer at pH 10. HQS complexes Cd2+ to form a fluorescent product. Thiols in the HPLC effluent compete for complexation of the Cd2+, resulting in a decrease in the fluorescence response. Detection limits of 0.2 microM (0.04 ppm) are achieved for cysteine, homocysteine and glutathione in a 5 min separation. Recoveries from spiked synthetic urine samples are 87-120%.Entities:
Mesh:
Substances:
Year: 2002 PMID: 12416880 DOI: 10.1016/s0021-9673(02)01087-7
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759