| Literature DB >> 12412683 |
A Schibler1, G L Hall, F Businger, B Reinmann, J H Wildhaber, M Cernelc, U Frey.
Abstract
Small airway disease in infants is characterised by abnormal lung volume and uneven ventilation distribution. An inert tracer gas washin/washout technique using a pulsed ultrasonic flow meter is presented to measure functional residual capacity (FRC) and ventilation distribution in spontaneously breathing and unsedated infants. With a pulsed ultrasound sent through the main stream of the flow meter, flow, volume and MM of the breathing gas can be calculated. Sulphur hexafluoride was used as a tracer gas. In a mechanical lung model (volume range 53-188 mL) and in 12 healthy infants (aged 38.3+/-9.2 days; mean+/-SD) accuracy and reproducibility of the technique was assessed. Indices of ventilation distribution such as alveolar-based mean dilution number (AMDN) and pulmonary clearance delay (PCD) were calculated. Mean error of volume measurement in the lung model was 0.58% (coefficient of variance (CV) 1.3%). FRC was in the low predicted range for normal infants (18.0+/-2.0 mL x kg(-1)) and highly reproducible (5.5+/-1.7% intra-subject CV). AMDN was 1.63+/-0.15 and PCD was 52.9+/-11.1%. Measurement of functional residual capacity and ventilation distribution using a sulphur hexafluoride washin/washout and an ultrasonic flow meter proved to be highly accurate and reproducible in a lung model and in healthy, spontaneously breathing and unsedated infants.Entities:
Mesh:
Year: 2002 PMID: 12412683 DOI: 10.1183/09031936.02.00226002
Source DB: PubMed Journal: Eur Respir J ISSN: 0903-1936 Impact factor: 16.671