Clinical performance of an in-line, ex vivo point-of-care monitor: a multicenter study.

Glenn F Billman1, Amy B Hughes, Golde G Dudell, Elizabeth Waldman, Lisa M Adcock, Dan M Hall, Edmund N Orsini, Adolph J Koska, Linda J Van Marter, Neil N Finer, Jeff C Kulhavy, Ronald D Feld, John A Widness.   

Abstract

BACKGROUND: The management of critically ill infants and neonates includes frequent determination of arterial blood gas, electrolyte, and hematocrit values. An objective of attached point-of-care patient monitoring is to provide clinically relevant data without the adverse consequences associated with serial phlebotomy.
METHODS: We prospectively determined the mean difference (and SD of the difference) from laboratory methods of an in-line, ex vivo monitor, the VIA LVM Blood Gas and Chemistry Monitoring System (VIA LVM Monitor; Metracor Technologies, Inc.), in 100 critically ill neonates and infants at seven children's hospitals. In doing so, we examined monitor stability with continuous use. In vivo patient test results from laboratory benchtop analyzers were compared with those from the VIA LVM Monitor on paired samples. In a separate in vitro comparison, benchtop analyzer and monitor test results were compared on whole-blood split samples.
RESULTS: A total of 1414 concurrent, paired-sample measurements were obtained. The mean differences (SD of differences) from laboratory methods and r values for the combined data for the VIA LVM Monitor from the seven sites were 0.001 (0.026) and 0.97 for pH, 0.7 (3.6) mmHg and 0.94 for PCO(2), 4.2 (9.6) mmHg and 0.98 for PO(2), 0.0 (2.9) mmol/L and 0.87 for sodium, 0.1 (0.2) mmol/L and 0.96 for potassium, and 0.3% (2.9%) and 0.90 for hematocrit. Performance results were similar among the study sites with increasing time of monitor use and between in vivo paired-sample and in vitro split-sample test results.
CONCLUSION: The VIA LVM Monitor can be used to assess critically ill neonates and infants.

Entities:  

Mesh:

Year:  2002        PMID: 12406990

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  2 in total

1.  Reduction in red blood cell transfusions among preterm infants: results of a randomized trial with an in-line blood gas and chemistry monitor.

Authors:  John A Widness; Ashima Madan; Ligia A Grindeanu; M Bridget Zimmerman; David K Wong; David K Stevenson
Journal:  Pediatrics       Date:  2005-05       Impact factor: 7.124

2.  Posttransfusion 24-hour recovery and subsequent survival of allogeneic red blood cells in the bloodstream of newborn infants.

Authors:  Ronald G Strauss; Donald M Mock; John A Widness; Karen Johnson; Gretchen Cress; Robert L Schmidt
Journal:  Transfusion       Date:  2004-06       Impact factor: 3.157

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.