Literature DB >> 12401621

The delivery rate accuracy of portable infusion pumps used for continuous regional analgesia.

Brian M Ilfeld1, Timothy E Morey, F Kayser Enneking.   

Abstract

UNLABELLED: Portable pumps used for local anesthetic infusion during continuous regional analgesia are gaining acceptance. These pumps are often used for ambulatory patients who are medically unsupervised throughout most of the infusion. However, the performance of these pumps, which infuse potentially toxic medication, has not been independently investigated. We investigated the flow rate accuracy, consistency, and profiles of various portable pumps often used for local anesthetic infusion during continuous regional analgesia. By using a computer/scale combination within a laboratory to record infusion rates, 6 pumps were tested with their flow regulators at expected (30 degrees-32 degrees C) and increased (34 degrees-36 degrees C) temperatures. Infusion rate accuracy differed significantly among the pumps, exhibiting flow rates within +/-15% of their expected rate for 18%-100% of their infusion duration. An increase in temperature also affected pumps to differing degrees, with infusion rates increasing from 0% to 25% for each model tested. These results suggest that factors such as flow rate accuracy and consistency, infusion profile, and temperature sensitivity should be considered when choosing and using a portable infusion pump for local anesthetic administration. IMPLICATIONS: Portable pumps often used for local anesthetic infusion during continuous regional analgesia exhibit varying degrees of delivery rate accuracy and consistency. Furthermore, increases in temperature result in an increased infusion rate for various pumps investigated. These factors should be taken into consideration when choosing and using a portable infusion pump.

Entities:  

Mesh:

Year:  2002        PMID: 12401621     DOI: 10.1097/00000539-200211000-00043

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  6 in total

1.  A randomized, triple-masked, active-controlled investigation of the relative effects of dose, concentration, and infusion rate for continuous popliteal-sciatic nerve blocks in volunteers.

Authors:  S J Madison; A M Monahan; R R Agarwal; T J Furnish; E J Mascha; Z Xu; M C Donohue; A C Morgan; B M Ilfeld
Journal:  Br J Anaesth       Date:  2014-09-23       Impact factor: 9.166

2.  Microfluidic Thermal Flowmeters for Drug Injection Monitoring.

Authors:  Il Doh; Daniel Sim; Steve S Kim
Journal:  Sensors (Basel)       Date:  2022-04-20       Impact factor: 3.847

Review 3.  Newer drug delivery systems in anesthesia.

Authors:  Sona Dave; Deepa Shriyan; Pinakin Gujjar
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2017 Apr-Jun

4.  Performance evaluation study of a commercially available smart patient-controlled analgesia pump with the microbalance method and an infusion analyzer.

Authors:  Jinsoo Park; Bongsu Jung
Journal:  J Dent Anesth Pain Med       Date:  2022-03-25

5.  The flow Rate Accuracy of Elastomeric Infusion Pumps After Repeated Filling.

Authors:  Masood Mohseni; Amin Ebneshahidi
Journal:  Anesth Pain Med       Date:  2014-04-07

6.  Protocol for a randomised crossover trial to evaluate patient and nurse satisfaction with electronic and elastomeric portable infusion pumps for the continuous administration of antibiotic therapy in the home: the Comparing Home Infusion Devices (CHID) study.

Authors:  Jodie G Hobbs; Melissa K Ryan; Brett Ritchie; Janet K Sluggett; Andrew J Sluggett; Lucy Ralton; Karen J Reynolds
Journal:  BMJ Open       Date:  2017-07-31       Impact factor: 2.692