Flat panel digital radiography compared with storage phosphor computed radiography: assessment of dose versus image quality in phantom studies.

Frank Fischbach1, Jens Ricke, Torsten Freund, Michael Werk, Birgit Spors, Clemens Baumann, Marcie J Pech, Roland Felix.   

Abstract

RATIONALE AND
OBJECTIVE: To assess and quantify the dose reduction by use of a CsI-flat panel digital radiography (DR)-system compared with digital computed radiography (CR).
MATERIALS AND METHODS: A TCDD-test using the CDRAD-phantom was performed at mAs-values of 5, 4, 2.5, 2, 1, and 0.5 mAs for both digital systems. Entrance surface doses were recorded for all images. Images were presented to four independent observers. For quantitative comparison the image quality figure (IQF) was calculated. Statistical analysis was performed using the Pearson correlation and the Wilcoxon test. A ROC analysis was performed using the TRG-phantom. Settings of 4, 2.5, 2 mAs for both systems were used. In addition, 1 and 0.5 mAs were used for the DR system only. Statistical significance was evaluated using Student test.
RESULTS: The DR system provided equivalent results compared with CR with respect to high frequency information and superior results with respect to low contrast details. Compared with computed radiography, the flat panel detector demonstrated significantly lower IQFs, ensuring a better image quality with respect to contrast and detail detectability. IQFs for DR and CR were equal at a surface dose reduction of 87% for DR. ROC analysis revealed significantly higher values under the curve for DR up to a surface dose reduction of 70%.
CONCLUSIONS: Image quality of DR proved to be far superior to CR in particular for low contrast details. The image quality of CR is similar to that of DR only at high dose levels.

Entities:  

Mesh:

Year:  2002        PMID: 12393973     DOI: 10.1097/00004424-200211000-00004

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  4 in total

1.  Emergency CT brain: preliminary interpretation with a tablet device: image quality and diagnostic performance of the Apple iPad.

Authors:  Patrick Mc Laughlin; Siobhan O Neill; Noel Fanning; Anne Marie Mc Garrigle; Owen J O Connor; Gerry Wyse; Michael M Maher
Journal:  Emerg Radiol       Date:  2011-12-16

2.  Optimisation of radiological protocols for chest imaging using computed radiography and flat-panel X-ray detectors.

Authors:  G Compagnone; M Casadio Baleni; E Di Nicola; M Valentino; M Benati; L F Calzolaio; N Oberhofer; E Fabbri; S Domenichelli; L Barozzi
Journal:  Radiol Med       Date:  2012-10-22       Impact factor: 3.469

3.  Introduction of a New Parameter for Evaluation of Digital Radiography System Performance.

Authors:  Mohammad Reza Choopani; Ali Chaparian
Journal:  J Med Signals Sens       Date:  2020-07-03

4.  CsI-detector-based dual-exposure dual energy in chest radiography for lung nodule detection: results of an international multicenter trial.

Authors:  Ricarda Rühl; Magdalena M Wozniak; Michael Werk; François Laurent; Georg Mager; Michel Montaudon; Andreas Pattermann; Antoine Scherrer; Jean-Pierre Tasu; Maciej Pech; Jens Ricke
Journal:  Eur Radiol       Date:  2008-04-19       Impact factor: 5.315

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.