Literature DB >> 12389103

A survey of radiation dose associated with pediatric plain-film chest X-ray examinations.

Kathleen M Hintenlang1, Jonathan L Williams, David E Hintenlang.   

Abstract

BACKGROUND: The effective doses delivered to pediatric patients are not well characterized for many plain-film X-ray examination techniques. The few data available on clinical doses in pediatric radiology are generally outdated because of the changes in X-ray generators and hardware that have taken place over the past decade.
OBJECTIVE: This survey characterizes X-ray examination techniques and measures effective doses delivered to a phantom representing a 1 year old in order to identify specific examination features that may result in greater than necessary doses to pediatric patients.
MATERIALS AND METHODS: An anthropomorphic phantom representing a 1 year old was developed for use as a survey tool. The phantom incorporates direct reading metal oxide semiconductor field effect transistor (MOSFET) dosimeters that permit the effective dose to be measured for clinical examinations. Seventeen commonly performed examinations were characterized at ten facilities with doses determined for a chest series of exams at each facility.
RESULTS: The survey demonstrates that the effective dose for a given examination can vary by an order of magnitude between institutions. Distributions of examination parameters identified those that are most significant for minimizing patient dose.
CONCLUSION: Efforts spent to determine pediatric specific radiographic techniques contribute more to effective imaging with low patient doses than utilizing AEC controls or high-frequency generators.

Entities:  

Mesh:

Year:  2002        PMID: 12389103     DOI: 10.1007/s00247-002-0734-3

Source DB:  PubMed          Journal:  Pediatr Radiol        ISSN: 0301-0449


  5 in total

Review 1.  Radiation safety.

Authors:  Donald P Frush
Journal:  Pediatr Radiol       Date:  2009-06

2.  Unnecessary irradiation to non-thoracic structures during pediatric chest radiography.

Authors:  Don Soboleski; Cynthia Theriault; Amy Acker; Vico Dagnone; David Manson
Journal:  Pediatr Radiol       Date:  2005-10-21

3.  Radiation exposure of image-guided intrathecal administration of nusinersen to adult patients with spinal muscular atrophy.

Authors:  D Oldenburg; N Guberina; B Stolte; K Kizina; E Stenzel; A Radbruch; C Kleinschnitz; T Hagenacker; M Forsting; C Mönninghoff
Journal:  Neuroradiology       Date:  2019-03-14       Impact factor: 2.804

4.  Revisit image control for pediatric chest radiography.

Authors:  Ehiichi Kohda; Yoshiyuki Tsutsumi; Masashi Nagamoto; Tatsuya Gomi; Hitoshi Terada; Yohko Kawawa; Hidekazu Masaki; Nobuyuki Shiraga
Journal:  Radiat Med       Date:  2007-02-27

5.  Estimation of radiation exposure of children undergoing superselective intra-arterial chemotherapy for retinoblastoma treatment: assessment of local diagnostic reference levels as a function of age, sex, and interventional success.

Authors:  Marcel Opitz; Denise Bos; Cornelius Deuschl; Alexander Radbruch; Sebastian Zensen; Selma Sirin; Michael Forsting; Nikolaos Bechrakis; Eva Biewald; Norbert Bornfeld; Petra Ketteler; Beate Timmermann; Martin Stuschke; Maja Guberina; Axel Wetter; Sophia Göricke; Nika Guberina
Journal:  Neuroradiology       Date:  2020-08-29       Impact factor: 2.804