| Literature DB >> 12385770 |
O Natt1, T Watanabe, S Boretius, J Radulovic, J Frahm, T Michaelis.
Abstract
This work demonstrates technical approaches to high-quality magnetic resonance imaging (MRI) of small structures of the mouse brain in vivo. It turns out that excellent soft-tissue contrast requires the reduction of partial volume effects by using 3D MRI at high (isotropic) resolution with linear voxel dimensions of about 100-150 microm. The long T(2)* relaxation times at relatively low magnetic fields (2.35 T) offer the benefit of a small receiver bandwidth (increased signal-to-noise) at a moderate echo time which together with the small voxel size avoids visual susceptibility artifacts. For measuring times of 1-1.5 h both T(1)-weighted (FLASH) and T(2)-weighted (Fast Spin-Echo) 3D MRI acquisitions exhibit detailed anatomical insights in accordance with histological sections from a mouse brain atlas. Preliminary applications address the identification of neuroanatomical variations in different mouse strains and the use of Mn(2+) as a T(1) contrast agent for neuroaxonal tracing of fiber tracts within the mouse visual pathway. Copyright 2002 Elsevier Science B.V.Entities:
Mesh:
Substances:
Year: 2002 PMID: 12385770 DOI: 10.1016/s0165-0270(02)00211-x
Source DB: PubMed Journal: J Neurosci Methods ISSN: 0165-0270 Impact factor: 2.390