Literature DB >> 12384743

Development and preclinical testing of a new tension-band device for the spine: the Loop system.

Matthew D Garner1, Steven J Wolfe, Stephen D Kuslich.   

Abstract

Wire sutures, cerclage constructs, and tension bands have been used for many years in orthopedic surgery. Spinous process and sublaminar wires and other strands or cables are used in the spine to re-establish stability of the posterior spinal ligament complex. Rigid monofilament wires often fail due to weakening created during twisting or wrapping. Stronger metal cables do not conform well to bony surfaces. Polyethylene cables have higher fatigue strength than metal cables. The Loop cable is a pliable, radiolucent, polyethylene braid. Creep of the Loop/locking clip construct is similar to metal cable constructs using crimps. Both systems have less creep than knotted polyethylene cable constructs.

Entities:  

Mesh:

Year:  2002        PMID: 12384743      PMCID: PMC3611575          DOI: 10.1007/s00586-002-0463-1

Source DB:  PubMed          Journal:  Eur Spine J        ISSN: 0940-6719            Impact factor:   3.134


  3 in total

1.  Proximal junctional failure after surgical instrumentation in adult spinal deformity: biomechanical assessment of proximal instrumentation stiffness.

Authors:  Maeva Lopez Poncelas; Luigi La Barbera; Jeremy Rawlinson; Dennis Crandall; Carl-Eric Aubin
Journal:  Spine Deform       Date:  2022-09-09

2.  Clinical experience of the dynamic stabilization system for the degenerative spine disease.

Authors:  Soo-Eon Lee; Sung-Bae Park; Tae-Ahn Jahng; Chun-Kee Chung; Hyun-Jib Kim
Journal:  J Korean Neurosurg Soc       Date:  2008-05-20

3.  Soft stabilization with an artificial intervertebral ligament in grade I degenerative spondylolisthesis: comparison with instrumented posterior lumbar interbody fusion.

Authors:  Chan Shik Shim; Sang-Ho Lee; Sun-Hee Park; Ji-Hee Whang
Journal:  SAS J       Date:  2007-08-01
  3 in total