Literature DB >> 12382809

Monte Carlo calculation and experimental verification of the photon energy response of tooth enamel in a head-sized plexiglas phantom.

A Wieser1, D Aragno, N El-Faramawy, P Fattibene, R Meckbach, S Onori, M C Pressello, L Pugliani, A Ulanovsky, M Zankl.   

Abstract

The use of electron paramagnetic resonance (EPR) tooth dosimetry for calculation of organ doses requires conversion of the measured absorbed dose in enamel. Before deriving conversion factors from simulation calculations with a realistic anthropomorphic human phantom, in the current study a simplified phantom was chosen to compare EPR measurement and Monte Carlo calculation. The dose response of tooth enamel of molars at various positions inside a cylindrical Plexiglas phantom of head-size was calculated hy Monte Carlo modelling in parallel photon beams of X rays of 63 keV equivalent energy and 60Co gamma rays (1.25 Mev). For X ray exposure, preliminary results of EPR dosimetry with tooth enamel samples prepared from molars irradiated in the phantom were in agreement with calculation. The mean value of the ratio of the measured to the calculated dose was 0.93 +/- 0.08.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12382809     DOI: 10.1093/oxfordjournals.rpd.a006046

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  1 in total

1.  Dosimetric response of tooth enamel to 14 Mev neutrons.

Authors:  P Fattibene; M Angelone; V De Coste; M Pillon
Journal:  Radiat Environ Biophys       Date:  2004-04-30       Impact factor: 1.925

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.