Literature DB >> 12382786

Experimental validation of Monte Carlo depth-dose calculations using radiochromic dye film dosimetry for a beta-gamma 153Sm radionuclide applied to the treatment of rheumatoid arthritis.

J E Villareal-Barajas1, G Ferro-Flores, O Hernandez-Oviedo.   

Abstract

In this work we compare the Monte Carlo (MCNP4B) calculated beta-gamma depth-dose profile for a liquid 153Sm beta-gamma source used in radiation synovectomy with the experimental depth-dose distribution obtained using radiochromic dye film dosimetry. The calculated and experimental depth-dose distribution shows a very good agreement (within 5%) in the region where the dose deposition is dominated by the beta particle component (first 800 microm depth on tissue-equivalent material). The agreement worsens, reaching a maximum deviation of 15%, at depths close to the maximum range of the beta particles. Finally the agreement improves for the region where the gamma component accounts for one-third of the total absorbed dose (depths >1 mm). The possible contributions to these differences are discussed, as well as their relevance for the application of 153Sm in the treatment of rheumatoid arthritis.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12382786     DOI: 10.1093/oxfordjournals.rpd.a006021

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  1 in total

Review 1.  Radiosynovectomy in the therapeutic management of arthritis.

Authors:  Liepe Knut
Journal:  World J Nucl Med       Date:  2015 Jan-Apr
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.