Clinical use of stereoscopic X-ray positioning of patients treated with conformal radiotherapy for prostate cancer.

Guy Soete1, Dirk Verellen, Dirk Michielsen, Vincent Vinh-Hung, Jan Van de Steene, Dirk Van den Berge, Patricia De Roover, Frans Keuppens, Guy Storme.   

Abstract

PURPOSE: To evaluate accuracy and time requirements of a stereoscopic X-ray-based positioning system in patients receiving conformal radiotherapy to the prostate. METHODS AND MATERIALS: Setup errors of the isocenter with regard to the bony pelvis were measured by means of orthogonal verification films and compared to conventional positioning (using skin drawings and lasers) and infrared marker (IR) based positioning in each of 261 treatments. In each direction, the random error represents the standard deviation and the systematic error the absolute value of the mean position. Time measurements were done in 75 treatments.
RESULTS: Random errors with the X-ray positioning system in the anteroposterior (AP), lateral, and longitudinal direction were (average +/- 1 standard deviation) 2 +/- 0.6 mm, 1.7 +/- 0.6 mm, and 2.4 +/- 0.7 mm. The corresponding values of conventional as well as IR positioning were significantly higher (p < 0.01). Systematic errors for X-ray positioning were 1.1 +/- 1.2 mm AP, 0.6 +/- 0.5 mm laterally, and 1.5 +/- 1.6 mm longitudinally. Conventional and IR marker-based positioning showed significantly larger systematic errors AP and laterally, but longitudinally, the difference was not significant. Depending on the axis looked at, errors of >or=5 mm occurred in 2%-14% of treatments after X-ray positioning, 13%-29% using IR markers, and 28%-53% with conventional positioning. Total linac time for one treatment session was 14 min 51 s +/- 4 min 18 s, half of which was used for the X-ray-assisted positioning procedure.
CONCLUSION: X-ray-assisted patient positioning significantly improves setup accuracy, at the cost of an increased treatment time.

Entities:  

Mesh:

Year:  2002        PMID: 12377349     DOI: 10.1016/s0360-3016(02)03027-4

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  8 in total

1.  Stereotactic body radiotherapy for prostate cancer: treatment approaches and clinical outcomes.

Authors:  Raymond Miralbell
Journal:  J Radiosurg SBRT       Date:  2011

2.  Residual translational and rotational errors after kV X-ray image-guided correction of prostate location using implanted fiducials.

Authors:  Reinhold Graf; Dirk Boehmer; Volker Budach; Peter Wust
Journal:  Strahlenther Onkol       Date:  2010-09-30       Impact factor: 3.621

3.  Incorporating imaging information from deep neural network layers into image guided radiation therapy (IGRT).

Authors:  Wei Zhao; Bin Han; Yong Yang; Mark Buyyounouski; Steven L Hancock; Hilary Bagshaw; Lei Xing
Journal:  Radiother Oncol       Date:  2019-07-11       Impact factor: 6.280

4.  Schedule for CT image guidance in treating prostate cancer with helical tomotherapy.

Authors:  G Beldjoudi; S Yartsev; G Bauman; J Battista; J Van Dyk
Journal:  Br J Radiol       Date:  2009-06-08       Impact factor: 3.039

5.  Is the in vivo dosimetry with the OneDosePlusTM system able to detect intra-fraction motion? A retrospective analysis of in vivo data from breast and prostate patients.

Authors:  Maria Daniela Falco; Marco D'Andrea; Alessia Lo Bosco; Mauro Rebuzzi; Elisabetta Ponti; Barbara Tolu; Grazia Tortorelli; Rosaria Barbarino; Luana Di Murro; Riccardo Santoni
Journal:  Radiat Oncol       Date:  2012-06-20       Impact factor: 3.481

6.  Optimal port placement and enhanced guidance in robotically assisted cardiac surgery.

Authors:  R Bauernschmitt; M Feuerstein; J Traub; Eva U Schirmbeck; G Klinker; R Lange
Journal:  Surg Endosc       Date:  2006-12-16       Impact factor: 3.453

7.  A novel compound 6D-offset simulating phantom and quality assurance program for stereotactic image-guided radiation therapy system.

Authors:  Dennis Yuen Kan Ngar; Michael Lok-Man Cheung; Michael Koon-Ming Kam; Wai-Sang Poon; Anthony Tak-Cheung Chan
Journal:  J Appl Clin Med Phys       Date:  2014-11-04       Impact factor: 2.102

8.  Performance and characteristics of an IR localizing system for radiation therapy.

Authors:  Yulia Lyatskaya; Hsiao-Ming Lu; Lee Chin
Journal:  J Appl Clin Med Phys       Date:  2006-05-25       Impact factor: 2.102

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.