Reconstruction of time-varying 3-D left-ventricular shape from multiview X-ray cineangiocardiograms.

Masamitsu Moriyama1, Yoshinobu Sato, Hiroaki Naito, Masayuki Hanayama, Takashi Ueguchi, Toshinobu Harada, Fujiichi Yoshimoto, Shinichi Tamura.   

Abstract

This paper reports on the clinical application of a system for recovering the time-varying three-dimensional (3-D) left-ventricular (LV) shape from multiview X-ray cineangiocardiograms. Considering that X-ray cineangiocardiography is still commonly employed in clinical cardiology and computational costs for 3-D recovery and visualization are rapidly decreasing, it is meaningful to develop a clinically applicable system for 3-D LV shape recovery from X-ray cineangiocardiograms. The system is based on a previously reported closed-surface method of shape recovery from two-dimensional occluding contours with multiple views. To apply the method to "real" LV cineangiocardiograms, user-interactive systems were implemented for preprocessing, including detection of LV contours, calibration of the imaging geometry, and setting of the LV model coordinate system. The results for three real LV angiographic image sequences are presented, two with fixed multiple views (using supplementary angiography) and one with rotating views. 3-D reconstructions utilizing different numbers of views were compared and evaluated in terms of contours manually traced by an experienced radiologist. The performance of the preprocesses was also evaluated, and the effects of variations in user-specified parameters on the final 3-D reconstruction results were shown to be sufficiently small. These experimental results demonstrate the potential usefulness of combining multiple views for 3-D recovery from "real" LV cineangiocardiograms.

Mesh:

Year:  2002        PMID: 12374315     DOI: 10.1109/TMI.2002.801161

Source DB:  PubMed          Journal:  IEEE Trans Med Imaging        ISSN: 0278-0062            Impact factor:   10.048


  3 in total

1.  Markov random field modeling for three-dimensional reconstruction of the left ventricle in cardiac angiography.

Authors:  Rubén Medina; Mireille Garreau; Javier Toro; Hervé L Breton; Jean-Louis Coatrieux; Diego Jugo
Journal:  IEEE Trans Med Imaging       Date:  2006-08       Impact factor: 10.048

2.  Evaluation of interpolation methods for surface-based motion compensated tomographic reconstruction for cardiac angiographic C-arm data.

Authors:  Kerstin Müller; Chris Schwemmer; Joachim Hornegger; Yefeng Zheng; Yang Wang; Günter Lauritsch; Christopher Rohkohl; Andreas K Maier; Carl Schultz; Rebecca Fahrig
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

3.  Interventional heart wall motion analysis with cardiac C-arm CT systems.

Authors:  Kerstin Müller; Andreas K Maier; Yefeng Zheng; Yang Wang; Günter Lauritsch; Chris Schwemmer; Christopher Rohkohl; Joachim Hornegger; Rebecca Fahrig
Journal:  Phys Med Biol       Date:  2014-04-15       Impact factor: 3.609

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.