Literature DB >> 12365020

Real-time tumor-tracking radiation therapy for lung carcinoma by the aid of insertion of a gold marker using bronchofiberscopy.

Toshiyuki Harada1, Hiroki Shirato, Shigeaki Ogura, Satoshi Oizumi, Koichi Yamazaki, Shinichi Shimizu, Rikiya Onimaru, Kazuo Miyasaka, Masaharu Nishimura, Hirotoshi Dosaka-Akita.   

Abstract

BACKGROUND: The authors developed fluoroscopic real-time tumor-tracking radiation therapy (RTRT) by insertion of a gold marker using bronchofiberscopy to reduce uncertainties in organ motion and set-up error in external radiotherapy for moving tumors. The purpose of the current study was to evaluate RTRT's feasibility in lung carcinoma treatment.
METHODS: The three-dimensional position of a 1.0-2.0 mm gold marker in or near the tumor was detected by two sets of fluoroscopies every 0.03 seconds. The treatment beam was gated to irradiate the tumor only when the position of the marker coincided with its planned position using the RTRT system. Bronchofiberscopic equipment for insertion of the marker into the lung tumor was developed and used for 20 lung tumors in 18 patients. Patients were given high dose hypofractionated focal irradiation (35-48 Gy in 4-8 fractions in 4-10 days) with a planning target volume margin of 5 mm for the tumor.
RESULTS: The markers were successfully inserted and maintained at the inserted position during and after the radiotherapy in 14 (88%) of 16 peripheral-type lung tumors and in none of four central-type lung tumors, indicating that this method of RTRT was not feasible for central-type lung tumors. Tracking of the marker was successfully performed in 1 of 2 tumors with a 1.0 mm marker and in all of 12 tumors with a 1.5-2.0 mm marker. On the whole, 13 (65%) of the 20 tumors were successfully treated with RTRT. Local tumor control was achieved and maintained for all 12 patients (13 tumors), who were treated with RTRT, with a median followup of 9 months (range, 5-15). Localized radiation pneumonitis was found radiographically at the lung volume that was irradiated with about 20 Gy, without symptoms in all but one patient.
CONCLUSIONS: The insertion of a gold marker into or near peripheral-type lung tumors using bronchofiberscopy is a feasible and safe technique. Excellent initial response and low incidence of clinical complications suggest that the high dose hypofractionated focal irradiation using the RTRT system can be a good local treatment for peripheral-type lung tumors. Copyright 2002 American Cancer Society.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12365020     DOI: 10.1002/cncr.10856

Source DB:  PubMed          Journal:  Cancer        ISSN: 0008-543X            Impact factor:   6.860


  18 in total

1.  [Ultrasound motion tracking for radiation therapy].

Authors:  J Jenne; J Schwaab
Journal:  Radiologe       Date:  2015-11       Impact factor: 0.635

Review 2.  Organ motion in image-guided radiotherapy: lessons from real-time tumor-tracking radiotherapy.

Authors:  Hiroki Shirato; Shinichi Shimizu; Kei Kitamura; Rikiya Onimaru
Journal:  Int J Clin Oncol       Date:  2007-02-25       Impact factor: 3.402

Review 3.  Techniques of endoscopic airway tumor treatment.

Authors:  Nicolas Guibert; Laurent Mhanna; Sylvain Droneau; Gavin Plat; Alain Didier; Julien Mazieres; Christophe Hermant
Journal:  J Thorac Dis       Date:  2016-11       Impact factor: 2.895

Review 4.  Improving radiotherapy planning, delivery accuracy, and normal tissue sparing using cutting edge technologies.

Authors:  Carri K Glide-Hurst; Indrin J Chetty
Journal:  J Thorac Dis       Date:  2014-04       Impact factor: 2.895

Review 5.  Fiducial marker placement for stereotactic body radiation therapy via convex probe endobronchial ultrasound: a case series and review of literature.

Authors:  Benjamin J Seides; John P Egan; Kim D French; Kevin L Kovitz; Neeraj R Desai
Journal:  J Thorac Dis       Date:  2018-03       Impact factor: 2.895

6.  Robust fluoroscopic tracking of fiducial markers: exploiting the spatial constraints.

Authors:  Rui Li; Gregory Sharp
Journal:  Phys Med Biol       Date:  2013-02-26       Impact factor: 3.609

7.  Analysis of fiducial markers used for on-line verification in the external-beam radiotherapy of patients with cranial tumours.

Authors:  R Ma Cañón; I Azinovic; M Lobato; J Navarro; J Rebollo
Journal:  Clin Transl Oncol       Date:  2007-08       Impact factor: 3.405

Review 8.  Navigational transbronchial needle aspiration, percutaneous needle aspiration and its future.

Authors:  Sixto Arias; Lonny Yarmus; A Christine Argento
Journal:  J Thorac Dis       Date:  2015-12       Impact factor: 2.895

Review 9.  Image-guided radiotherapy and motion management in lung cancer.

Authors:  S S Korreman
Journal:  Br J Radiol       Date:  2015-05-08       Impact factor: 3.039

10.  Magnitude of shift of tumor position as a function of moderated deep inspiration breath-hold: An analysis of pooled data of lung patients with active breath control in image-guided radiotherapy.

Authors:  K R Muralidhar; P Narayana Murthy; D Shankar Mahadev; K Subramanyam; G Sudarshan; A Krishnam Raju
Journal:  J Med Phys       Date:  2008-10