Dunmin Mao1, Chuanfan Ding, D J Douglas.
Abstract
The hydrogen/deuterium (H/D) exchange of gas-phase ions of holo- and apo-myoglobin has been studied by confining the ions in a linear quadrupole ion trap with D(2)O or CD(3)OD at a pressure of several mTorr. Apo-myoglobin ions were formed by collision-induced dissociation of holo-myoglobin ions between the orifice and skimmer of the ion sampling system. The exchange takes place on a time scale of seconds. Earlier cross section measurements have shown that holo-myoglobin ions can have more compact structures than apo-myoglobin. Despite this, both holo-myoglobin and apo-myoglobin in charge states +8 to +14 are found to exchange nearly the same number of hydrogens (ca. 103) in 4 s. It is possible the ions fold or unfold to new conformations on the much longer time scale of the exchange experiment compared with the cross section measurements. Copyright 2002 John Wiley & Sons, Ltd.Entities:
Mesh:
Substances:
Year: 2002 PMID: 12362385 DOI: 10.1002/rcm.810